CONFIGURATION: Setup For Use
Revision A
DASH 2000 Patient Monitor
6-25
2000 412-001
3.
The SET UNIT NAME pop-up window appears on the monitor
display as shown below.
The SET UNIT NAME pop-up window displays either the current
care UNIT NAME or is completely blank. The software supports up
to seven alphanumeric characters to be used in the UNIT NAME
field.
NOTE
It is important that you enter the correct UNIT NAME
with regard to spelling, spaces and special characters
programmed into the field. If a mistake is made in
programming the UNIT NAME field, the monitor is not
available on the network for display at central stations
within the same care unit.
4.
Two sets of arrow icons appear in the UNIT NAME pop-up window.
¥
The horizontal (left/right) arrow icons, when highlighted, allow
the user to select a specific character for change by rotating the
Trim Knob control. Press and release of the Trim Knob while a
specific character is highlighted enables that specified character
for change.
¥
The vertical (up/down) arrow icons, when highlighted, allow the
user to scroll through all of the alpha numerics available for each
character. Rotating the Trim Knob control at this point allows the
user to select a specific alphanumeric to be entered into the
specified character position within this field. To enter the chosen
character into memory, press and release the Trim Knob control
one more time.
5.
Repeat step 4 for each character to be entered as part of the UNIT
NAME. Up to seven characters may be setup or configured. The
manufacturer recommends UNIT NAME fields that are less than
seven characters to be left-justified, leaving unused character
positions (immediately to the right of the user-entered unit name)
blank.
6.
When you finish making each character entry, use the Trim Knob
control to select the SET UNIT NAME menu option item. Press the
Trim Knob control. This programs the newly entered UNIT NAME
into the monitor memory and closes the pop-up window.
MAIN
MENU
PREV.
MENU
SET
UNIT
NAME
Содержание marquette Dash 2000
Страница 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Страница 159: ......
Страница 173: ......
Страница 178: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 184: ......
Страница 196: ......
Страница 197: ......
Страница 198: ......
Страница 200: ......
Страница 201: ......