Maintenance: Checkout Procedures
Revision A
DASH 2000 Patient Monitor
3-11
2000 412-001
6.
Verify accuracy of the SPO
2
% values (these are the white NELLCOR
values shown on the SpO
2
simulator) on the monitor display using
the SpO
2
simulator settings from the following table:
7.
Verify accuracy of the PPR values on the monitor display using the
SpO
2
simulator pulse rates from the following table.
8.
Press the INTERFERENCE TEST button on the SpO
2
simulator for
30 seconds.
9.
Verify the displayed SPO
2
% value remains 92 Ð 99%, or an
interference detection message is displayed and XX is displayed in
the SpO
2
parameter window in place of an SPO
2
% value.
10. Set the SpO
2
simulator power switch to the off position.
11. Disconnect the Nellcor-style SpO
2
simulator cable from the monitor
SpO
2
connector. This completes the SpO
2
tests.
Noninvasive Blood
Pressure Test
The overall accuracy of noninvasive blood pressure (NBP) readings by
the monitor depend on the following:
¥
the zero pressure reading, and
¥
the voltage span of the NBP sensor in the monitor.
This procedure provides a method of verifying these items are accurate
and also checks the NBP pneumatic circuit plumbing for leaks.
WARNING
When the NBP cuff is used in this procedure, it must be
tightly wrapped around a rigid cylinder or pipe.
Do not
put the NBP cuff around a human arm during the
calibration procedures due to the potential for injury.
1.
Remove all cables except for the power cord from the monitor.
SpO
2
Simulator Setting
Displayed SPO
2
% Value
95.5%
92 Ð 99
85.5%
82 Ð 89
68.4%
65 Ð 72
Simulator PULSE RATE
Displayed PPR Value
70 B/M
66 Ð 74
100 B/M
96 Ð 104
160 B/M
155 Ð 165
Содержание marquette Dash 2000
Страница 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Страница 159: ......
Страница 173: ......
Страница 178: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 184: ......
Страница 196: ......
Страница 197: ......
Страница 198: ......
Страница 200: ......
Страница 201: ......