Maintenance: Checkout Procedures
3-14
DASH 2000 Patient Monitor
Revision A
2000 412-001
8.
Rotate the Trim Knob control to highlight CALIBRATE AND TEST
and press the Trim Knob control to select it. Next, rotate the Trim
Knob control to highlight CALIBRATE NBP and press the Trim
Knob control to select it.
9.
Rotate the Trim Knob control to highlight CHECK CAL OFF, and
then press the Trim Knob control to select it.
10. Rotate the Trim Knob control to highlight START, and then press the
Trim Knob control to select it.
11. Leakage Test
The Dash pumps up to 250 mmHg and then holds the pressure. Wait
about 30 s until the pressure has stabilized to approximately
240 mmHg. From now on the dropdown-rate of the pressure must be
less then 4 mmHg/min.
12. Measurement Accuracy
By means of the latex bulb adjust the following pressures and check
that the tolerance limits are not exceeded.
250 mmHg
±5 mmHg
200 mmHg
±4 mmHg
150 mmHg
±3 mmHg
100 mmHg
±3 mmHg
50 mmHg
±3 mmHg
13. Deflation Pressure Threshold Test
Increase the pressure, the cuff must be deflated automatically
between 300 mmHg and 330 mmHg.
Leave the service mode by pressing MAIN MENU:
In NIBP menu, select neonate cuff size. Now start NIBP
measurement. The system pumps up to appr. 120 mmHg. Now
increase the pressure with the bulb, the cuff must be deflated
automatically between 150 mmHg and 165 mmHg.
14. Active Test
Apply a cuff and measure the blood pressure. Asses that the SYS,
MAP and DIA parameter readings are plausible.
MAIN
MENU
MONITOR /
ADMIT
MODE
TIME
AND
DATE
BATTERY
SERVICE
CALIBRATE
AND
TEST
ADDRESS
REVIEW
ERRORS
MAIN
MENU
GRAPH
TEST
PATTERN
CALIBRATE
NBP
CO2
SERVICE
PREV.
MENU
MAIN
MENU
CAL
ZERO:
OFF
CHECK
CAL:
OFF
PREV.
MENU
CAL
GAIN:
OFF
Содержание marquette Dash 2000
Страница 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Страница 159: ......
Страница 173: ......
Страница 178: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 184: ......
Страница 196: ......
Страница 197: ......
Страница 198: ......
Страница 200: ......
Страница 201: ......