CALIBRATION: Noninvasive Blood Pressure
Revision A
DASH 2000 Patient Monitor
5-7
2000 412-001
4.
Connect a cuff and manometer to the monitor as shown below.
5.
Turn the manometer on and adjust the range switch to the
1000ÊmmHg setting.
NBP tubing
pn 414873-001
Coupling
pn 46100-002:
Connects manometer
tubing to NBP tubing.
NBP cuff
pn 9461-301:
Any size NBP cuff will
be sufficient.
6-inch diameter
PVC pipe (or
1-pound coffee can):
Wrap the NBP cuff
around this for tests.
3-way tee
pn 4745-101:
Connects manometer
tubing, NBP tubing
and NBP cuff tubing.
Digital manometer
Sensym PDM200M:
Note: A calibrated
mercury manometer
may be substituted.
Manometer tubing
pn 401582-001:
Approximately two
feet in length.