CONFIGURATION: Load Software From Diskette
Revision A
DASH 2000 Patient Monitor
6-11
2000 412-001
Errors During Download
Process
For most errors, simply press RETURN on the PC or repeat the
download procedure. If the byte numbers stop advancing for more than
two minutes, refer to ÒProblems While Loading SoftwareÓ found in the
ÒIntroductionÓ section of this procedure.
Repeat Steps For Each
File Requiring Update
Perform steps in ÒSetup Monitor to Accept Download FilesÓ and
ÒDownload Files to the MonitorÓ for each file that requires updating
before proceeding to the next steps. When all code is loaded, turn monitor
power off, then on.
Completion
To configure (setup) proper graph locations for a monitor that has a
standalone writer attached to it, connect the monitor to the network then
follow these steps:
Setup Graph Locations
Use the Trim Knob to scroll to and select the following menu options
selection sequence. Beginning at the monitor main menu options, scroll
to and select:
MONITOR SETUP
GRAPH SETUP
GRAPH LOCATION
Select a Writer
From the GRAPH LOCATION menu option items, use the Trim Knob to
scroll to and select:
¥
the MANUAL GRAPH LOCATION option item (the monitor may
take up to a minute to poll the network for available writers), then
choose one of the manual graph locations from that list of writers;
then
¥
the ALARM GRAPH LOCATION option item, then choose one of the
alarm graph locations from that list of writers; and finally
¥
the PRINT WINDOW LOCATION option item, then choose one of the
print window locations from that list of writers.
Test the Monitor
Connect a patient simulator to the monitor. Admit and generate patient
waveforms at the monitor with the simulator powered up. Perform the
following steps to test the communication paths between the monitor and
each selected writer.
¥
Press the GRAPH GO/STOP key on the monitor front panel. Verify
the graph output arrives at the selected manual graph location.
Press the GRAPH GO/STOP key again to stop the manual graph.
¥
Switch the simulator power off to cause a fatal alarm by. Verify the
graph output arrives at the selected alarm graph location.
¥
Bring up a non-real-time window on the monitor display. Print the
window. Verify the print output arrives at the selected print window
location.
Содержание marquette Dash 2000
Страница 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Страница 159: ......
Страница 173: ......
Страница 178: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 184: ......
Страница 196: ......
Страница 197: ......
Страница 198: ......
Страница 200: ......
Страница 201: ......