CONFIGURATION: Load Software Over The Network
6-16
DASH 2000 Patient Monitor
Revision A
2000 412-001
5.
From the SCRIPT NAME SELECTION menu list, use the Trim Knob
to scroll to and select the number corresponding to the file (script)
requiring update. Following is a list of files that appear in the
SCRIPT NAME SELECTION menu (listed in sequential order for
each script to be loaded):
*MAIN.SCR,
*BOOT.SCR,
DASMAIN.SCR, and
DASBOOT.SCR.
*Monitor reboots when loading these scripts.
Download only the files that require update based on comparison of
file revisions made previously. Once a file has been selected, the
monitor will begin the download process.
6.
The monitor should display the part number, version, and date of the
file to be downloaded.
NOTE
A warning message and prompt appears on the monitor
display. Use the Trim Knob to scroll to and select YES if
the file selected for download is correct.
CAUTION
Do not
reboot or power down the monitor while
downloading boot code components (BOOT.SCR,
DASBOOT.SCR, etc.). This renders the monitor useless
and manufacturer factory service is required.
7.
The order in which the files are updated in the monitor is important
(see ÒNetwork Download ProcedureÓ). If an update of the main
processor operational code (MAIN.SCR) or main processor boot code
(BOOT.SCR) components is required, the monitor reboots
automatically upon completion of each of those updates.
8.
Messages appear on the display to indicate how the update is going.
9.
For most errors, simply repeat the previous steps. If the byte
numbers stop advancing for more than two minutes, start the
procedure over or call technical support.
10. Perform the previous steps for each software file as required. This
should be based on comparison of revisions made earlier in this
procedure.
11. When the update is complete, use the Trim Knob to select START
PATIENT MONITORING.
Содержание marquette Dash 2000
Страница 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Страница 159: ......
Страница 173: ......
Страница 178: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 184: ......
Страница 196: ......
Страница 197: ......
Страница 198: ......
Страница 200: ......
Страница 201: ......