CALIBRATION: Noninvasive Blood Pressure
5-8
DASH 2000 Patient Monitor
Revision A
2000 412-001
Start the Gain Calibration
Test
1.
Rotate the Trim Knob control to highlight CAL GAIN OFF, and then
press the Trim Knob control to select it.
2.
Rotate the Trim Knob control to highlight CAL GAIN OFF, and then
press the Trim Knob control to select it.
3.
Rotate the Trim Knob control to highlight START, and then press the
Trim Knob control to select it. The second line of text on the CAL
GAIN menu item changes from OFF to INFLATING. Then, the
monitor starts pumping up the pressure bulb or cuff Ð the audible
whirring sound of the NBP pump motors occurs and an increase in
displayed pressures on both the monitor and the manometer can be
observed.
4.
The pump shuts off at about 250 mmHg, and the pressure drops
slowly to about 240 mmHg before stabilizing. The second line of text
on the CAL GAIN menu item changes from INFLATING back to
HOLDING. If the pressure continues to drop at a rate of
4 mmHg/min or more, there is a leak in the NBP plumbing. If there
is a leak in the NBP plumbing, repair it and restart this calibration
procedure.
MAIN
MENU
CAL
ZERO:
OFF
CHECK
CAL:
OFF
PREV.
MENU
CAL
GAIN:
OFF
MAIN
MENU
CAL
GAIN:
OFF
PREV.
MENU
ENTER
CAL
PRESSURE
MAIN
MENU
>
START
>
STOP
CAL
GAIN:
OFF
PREV.
MENU
MAIN
MENU
CAL
GAIN:
INFLATING
PREV.
MENU
ENTER
CAL
PRESSURE
MAIN
MENU
CAL
GAIN:
INFLATING
PREV.
MENU
ENTER
CAL
PRESSURE
MAIN
MENU
CAL
GAIN:
HOLDING
PREV.
MENU
ENTER
CAL
PRESSURE
Содержание marquette Dash 2000
Страница 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Страница 159: ......
Страница 173: ......
Страница 178: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 184: ......
Страница 196: ......
Страница 197: ......
Страница 198: ......
Страница 200: ......
Страница 201: ......