Maintenance: Checkout Procedures
Revision A
DASH 2000 Patient Monitor
3-7
2000 412-001
NOTE
If indicator blinks yellow, then there is a malfunction in
the power management system. The unit needs to be
repaired.
4.
Switch the unit on and disconnect the the power cord. Verify that the
BATTERY indicator stays illuminated, AC power indicator and
CHARGING STATUS indicator go off.
5.
Connect the power cord back to the unit. Before continuing the test
procedures the battery condition has to be checked.
Starting from the screen showing the softbutton MORE MENUS,
select the following menus in sequence:
MORE MENUS, MONITOR SETUP, SERVICE MODE (actual date
as password), BATTERY SERVICE.
If the entry for last conditioning is older then 3 months, start a
condition cycle (takes hours). If the entry is within 3 months, check if
the FULL/NEW entry is below 40%, if it is below, exchange the
battery.
6.
Run the unit under battery power and perform the ECG Test step 1.
to 4. as described later.
7.
Connect the unit back to AC and perform the following tests.
ECG Tests
1.
Set up the patient simulator as follows:
¥
Heart rate Ð 80 bpm,
¥
Heart rate amplitude Ð 1.0 mV,
¥
5-lead ECG patient cable properly attached.
2.
Attach the ECG patient cable and ECG leadwire set to the ECG/
RESP connector on the monitor and the leadwire connectors on the
top of the patient simulator.
3.
Admit the patient simulator to the monitor.
4.
Observe the following:
¥
ECG lead II is displayed and is noise-free,
¥
Heart rate of 80 ±2 bpm is displayed,
¥
With QRS tones enabled, an audible tone sounds with each R-
Wave (QRS complex).
5.
Verify all seven ECG leads are available for viewing and are noise-
free.
6.
Select ANALYSIS SETTING, DETECT PACE and set to PACE2.
7.
Set ECG amplitude on simulator to 2 mV. Select the VP2 pacemaker
pulse on the simulator.
Содержание marquette Dash 2000
Страница 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Страница 159: ......
Страница 173: ......
Страница 178: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 184: ......
Страница 196: ......
Страница 197: ......
Страница 198: ......
Страница 200: ......
Страница 201: ......