Maintenance: Checkout Procedures
Revision A
DASH 2000 Patient Monitor
3-9
2000 412-001
Respiration Tests
1.
With the ECG patient cable still connected to the ECG/RESP
connector of the monitor, set up the patient simulator as follows:
¥
Respiration (RESP) baseline impedance Ð 750 ½,
¥
RESP ÆR Ð 0.5 ½,
¥
RESP lead select Ð I & II,
¥
RESP rate (respirations per minute) Ð 30.
2.
Set up the monitor as follows:
¥
RESP parameter Ð on
¥
RESP waveform Ð on,
¥
RESP waveform lead select Ð lead II (RESP waveform derived
from ECG lead II).
3.
Observe the following:
¥
RESP parameter window appears on the monitor with a reading
of 30 ±2 (respirations per minute),
¥
RESP waveform appears distortion-free on the monitor.
4.
Change the RESP waveform lead select of the monitor to lead I
(RESP waveform derived from ECG lead I).
5.
Observe the following:
¥
RESP parameter window appears on the monitor with a reading
of 30 ±2 (respirations per minute),
¥
RESP waveform appears distortion-free on the monitor.
6.
Disconnect the ECG patient cable from the ECG/RESP connector of
the monitor. Proceed to the next steps in these checkout procedures.
Temperature Tests
1.
Set up the patient simulator for a temperature output of 37 ¡C.
2.
Attach the temperature simulator from the series 400
TEMPERATURE OUTPUT connector of the patient simulator to the
TEMP input of the monitor.
3.
Verify a TEMP parameter window appears on the monitor display
with a temperature reading of 37.0 ¡C ±0.4 ¡C.
4.
Remove the temperature adaptor and temperature simulator cable
from the monitor and patient simulator.
Invasive Blood
Pressure Tests
The invasive blood pressure (BP) tests provide a method of verification
for the BP connector of a monitor equipped with this optional function.
Follow these steps:
1.
Set up the patient simulator as follows:
¥
Blood pressure (BP) polarity Ð POS,
¥
BP output Ð 0 mmHg.
Содержание marquette Dash 2000
Страница 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Страница 159: ......
Страница 173: ......
Страница 178: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 184: ......
Страница 196: ......
Страница 197: ......
Страница 198: ......
Страница 200: ......
Страница 201: ......