CALIBRATION: Adjustments and Switches
5-2
DASH 2000 Patient Monitor
Revision A
2000 412-001
Adjustments and Switches
Hardware Calibration
The following table summarizes the hardware adjustments and switches
on the monitor. The hardware adjustments are only necessary if a circuit
board is repaired or replaced.
Software Calibration
Noninvasive blood pressure (NBP) is the only function that requires
software calibration. The manufacturer recommends performing the
NBP software calibration upon receipt of the monitor initially, and once
each year thereafter. The NBP software calibration should also be
performed whenever the monitor is opened for service purposes. This
ensures the pneumatic circuit plumbing has not developed any air leaks
as a result of disassembly.
Reference
Designation
Main Processor PCB:
S1
2-station DIP switch, LCD Type and Watchdog
Main Connector PCB:
None
None
Acquisition PCB:
None
None
Power Supply PCB:
None
None
Содержание marquette Dash 2000
Страница 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Страница 159: ......
Страница 173: ......
Страница 178: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 184: ......
Страница 196: ......
Страница 197: ......
Страница 198: ......
Страница 200: ......
Страница 201: ......