CALIBRATION: Noninvasive Blood Pressure
5-10
DASH 2000 Patient Monitor
Revision A
2000 412-001
as CUFF in the NBP parameter box) on the monitor and manometer
are equal (± 1 mmHg) for
at least
one full minute.
11. Rotate the Trim Knob control to highlight CHECK CAL IN
PROGRESS and press the Trim Knob control to select it.
12. Rotate the Trim Knob control to highlight STOP and press the Trim
Knob control to select it. The monitor automatically releases
pneumatic pressure in the entire plumbing circuit.
13. Turn the monitor off, turn the manometer off and remove the test
apparatus from the monitor.
ECG
150/ 50
21-JUN-1999 16:27
II
V5
I
ICU-BED 5
MARTIN
X
NBP
200/ 80
S
CUFF
X
250
X /
MAIN
MENU
PREV.
MENU
CAL
ZERO:
OFF
CHECK
CAL:
IN PROG
CAL
GAIN:
OFF
MAIN
MENU
CAL
ZERO:
OFF
CHECK
CAL:
IN PROG.
PREV.
MENU
CAL
GAIN:
OFF
>START
>
STOP
CHECK
CAL:
IN PROG.
Содержание marquette Dash 2000
Страница 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Страница 159: ......
Страница 173: ......
Страница 178: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 184: ......
Страница 196: ......
Страница 197: ......
Страница 198: ......
Страница 200: ......
Страница 201: ......