Maintenance: Checkout Procedures
Revision A
DASH 2000 Patient Monitor
3-13
2000 412-001
6.
Connect a cuff and manometer to the monitor as shown below.
7.
The service menus should appear on the monitor display. These next
steps guide you through the service menus associated with checking
NBP calibration and checking for leaks. If desired test results are not
obtained, NBP calibration is necessary.
NBP tubing
pn 414873-001
Coupling
pn 46100-002:
Connects manometer
tubing to NBP tubing.
NBP cuff
pn 9461-301:
Any size NBP cuff will
be sufficient.
6-inch diameter PVC
pipe (or
1-pound coffee can):
Wrap the NBP cuff
around this for tests.
3-way tee
pn 4745-101:
Connects manometer
tubing, NBP tubing
and NBP cuff tubing.
Digital manometer
Sensym PDM200M:
Note: A calibrated
mercury manometer
may be substituted.
Manometer tubing
pn 401582-001:
Approximately two
feet in length.
Coupling
pn 400787-001:
Connects manometer
tubing to NBP cuff
tubing.
Latex bulb
with deflation valve
BAUM
pn 0661-2941
Содержание marquette Dash 2000
Страница 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Страница 159: ......
Страница 173: ......
Страница 178: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 184: ......
Страница 196: ......
Страница 197: ......
Страница 198: ......
Страница 200: ......
Страница 201: ......