CALIBRATION: Noninvasive Blood Pressure
Revision A
DASH 2000 Patient Monitor
5-9
2000 412-001
5.
Rotate the Trim Knob control to highlight ENTER CAL PRESSURE
and press the Trim Knob control to select it.
6.
An ENTER CAL PRESSURE pop-up window will appear. Use the
Trim Knob control to select a pressure value that is 1 mmHg lower
than the current manometer reading.
7.
When the manometer falls to exactly the value that you selected in
the pop-up window, press the Trim Knob control to enter the value.
8.
Rotate the Trim Knob control PREV. MENU to highlight CHECK
CAL OFF, and then press the Trim Knob control to select it.
9.
Rotate the Trim Knob control to highlight START and press the Trim
Knob control to select it.
10. The text on the menu item changes from CHECK CAL OFF to
CHECK CAL IN PROGRESS. Verify the pressure readings (shown
MAIN
MENU
CAL
GAIN:
HOLDING
PREV.
MENU
ENTER
CAL
PRESSURE
MAIN
MENU
250
ENTER
CAL
PRESSURE
PREV.
MENU
MAIN
MENU
CAL
ZERO:
OFF
CHECK
CAL:
OFF
PREV.
MENU
CAL
GAIN:
OFF
MAIN
MENU
>
START
>
STOP
CAL
ZERO:
OFF
CHECK
CAL:
OFF
PREV.
MENU
CAL
GAIN:
OFF
Содержание marquette Dash 2000
Страница 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Страница 159: ......
Страница 173: ......
Страница 178: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 184: ......
Страница 196: ......
Страница 197: ......
Страница 198: ......
Страница 200: ......
Страница 201: ......