CONFIGURATION: Load Software From Diskette
Revision A
DASH 2000 Patient Monitor
6-7
2000 412-001
Update Program
Start-up
Start the update download program from an update diskette to begin
loading software into the monitor by following these steps:
1.
Apply power to the PC and wait for the
C:\>
prompt to appear on the
PC display.
CAUTION
The manufacturer recommends operating the PC (or PC
laptop) on AC power for the duration of the update
process. This prevents inadvertent power interruptions
to the PC or PC laptop. Interruptions of power cause the
update process to fail. While downloading the boot code
components, interruptions in the update process may
result in monitor malfunction or being rendered
completely useless. The monitor may require factory
service as a result.
NOTE
If the PC used for this procedure automatically launches
any version of Windows, perform the necessary steps to
quit Windows and return to DOS.
2.
Compare the SOFTWARE SUMMARY window with the file names
from the tables on the next page. Only load the files that currently
reflect earlier revisions, as compared with the SOFTWARE
SUMMARY printout, into the monitor. Generally, the main processor
operational code (MAIN) or acquisition processor operational code
(DAS) need to be updated. Depending on the vintage of the monitor,
boot code may need to be updated as well but this is generally not the
case.
3.
Following is the order in which the update files are to be downloaded:
¥
Main processor op-code (MAIN), and
¥
Acquisition processor op-code (DAS).
Then, only if necessary:
¥
Main processor boot code (MAIN BOOT), and
¥
Acquisition processor boot code (DAS BOOT).
Содержание marquette Dash 2000
Страница 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Страница 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Страница 159: ......
Страница 173: ......
Страница 178: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 184: ......
Страница 196: ......
Страница 197: ......
Страница 198: ......
Страница 200: ......
Страница 201: ......