14
●
EMC Guidance
This device complies with the Safety Standard IEC 60601-1-2 (2004). However, if portable transmitter or wireless LAN
equipment is used extremely nearby, the electromagnetic influence may largely exceed the compliance level and may cause
unexpected phenomenon such as noise interference on the waveform, etc.
This device should be used in a location specified by each medical institution.
If any unexpected noise interference on the waveform or failure to the peripheral device occurs, stop using the device and follow
the instruction of the technical engineer.
The following is the information relating to EMC (Electromagnetic Compatibility).
(When using this device, verify that it is used within the environment specified below.)
●
Compliance to the Electromagnetic Emissions
The VS-2000 is intended for use in the electromagnetic environment specified below.
Emission Test
Compliance
Electromagnetic Environment/Guidance
RF emissions:
CISPR 11
Group 1
The VS-2000 uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions:
CISPR 11
Class B
The VS-2000 is suitable for use in all establishments, including domestic, and
those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Содержание VaSera VS-2000 Series
Страница 4: ......
Страница 22: ...18 ...
Страница 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Страница 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Страница 172: ...9 22 Chapter 9 File Transfers Communication History ...
Страница 176: ...10 4 Chapter 10 Recording Daily Reports ...
Страница 214: ...12 12 Chapter 12 Maintenance and Inspection ...
Страница 233: ...The company and product names used in this manual are trademarks or registered trademarks of respective companies ...