11-17
Chapter 11 Settings of This Device
Printer
Settings for Blood Pressure and Pulse Wave
Examinations
Pulse Wave: General
File
: Indicates settings that are displayed in Simple Examination Mode.
Setup Item
Description
Setting Range
<Network Printer>
Network Printer
Specifies whether a printer is connected via the wired LAN or wireless LAN. ON, [OFF]
IP Address
If "Network Printer" is set to "ON", specify the IP address of the printer.
[0.0.0.0]
Port No.
Specifies the port number.
[9100]
Maximum 5 digits
<PC Printer>
PC Printer
Specifies whether a printer is connected via the wired LAN or wireless LAN. ON, [OFF]
IP Address
If "PC Printer" is set to "ON", specify the IP address of the printer.
[0.0.0.0]
Port No.
Specifies the port number.
[18200]
Maximum 5 digits
Setup Item
Description
Setting Range
Anonymous
Specifies whether to make patient information anonymous.
ON: The name, age and gender are not displayed or printed.
OFF: The information is not made anonymous.
ON, [OFF]
Buzzer
Specifies the buzzer setting.
[Buzzer Control 1],
Buzzer Control 2,
Buzzer Control 3
Confirm results
Specifies whether to confirm the result display before performing actions such
as saving after taking measurements.
ON, [OFF]
Default Measurement Site Set the default measurement site. The site set here is set as the measurement
site when the power is turned on or when switching to a new patient.
Multiple items can be selected.
[RB], [LB],
[RA], [LA],
RT, LT
Default Measurements
Set the default measurement item. The item set here is set as the measurement
item when the power is turned on or when switching to a new patient.
Multiple items can be selected.
[CAVI/Pulse Wave],
[ABI/Limb BP],
TBI
Setup Item
Description
Setting Range
Auto Save
Specifies whether to automatically save blood pressure and pulse wave data
after taking measurements.
This is enabled when [File] - [General] - [Auto Save] is set to "Each Exam. Mode".
ON, [OFF]
Storage Media
Specifies the media to save the data.
Multiple items can be specified.
[SD Card],
USB1,2,3, DMS
Trend Count
Specifies the maximum number of trends to be printed in reports.
10, [50]
Compatible with VS-1500
Series
Saves data in the same format as the VS-1500 Series.
ON, [OFF]
Compare Waveforms
Specifies whether to automatically obtain data when conducting trend examinations. ON, [OFF]
T
T
Содержание VaSera VS-2000 Series
Страница 4: ......
Страница 22: ...18 ...
Страница 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Страница 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Страница 172: ...9 22 Chapter 9 File Transfers Communication History ...
Страница 176: ...10 4 Chapter 10 Recording Daily Reports ...
Страница 214: ...12 12 Chapter 12 Maintenance and Inspection ...
Страница 233: ...The company and product names used in this manual are trademarks or registered trademarks of respective companies ...