2-16
Chapter 2 Installation
Connecting to the Network
Follow the steps below to connect to the network to send examination results from the device to a computer.
Example of Network Configuration
Connect the devices as shown in the figure below.
An Ethernet hub and LAN cable are required. Connect the LAN cable to the LAN connector of the main unit and connect the
other end to the Ethernet hub. Connect the PC in the same way.
Required Items
• Ethernet hub
• LAN cable (straight) x 2
If the settings need to be changed according to an existing network environment, refer to "Network Settings (Wired LAN)" (P2-
17) and change the settings.
• The devices can also be connected by wireless LAN. If connecting by wireless LAN, factors such as obstructions may prevent
correct communication in some installation locations. Check that correct communication is possible when installing the device.
• Configure and connect to the network as specified by Fukuda Denshi.
• Use an external printer and Ethernet hub specified by Fukuda Denshi.
• The devices connected to the network, such as the external printer, Ethernet hub and computer,
need to be installed 1.5m away from the patient (IEC 60601-1-1: 2000/EN 60601-1-1: 2001).
• The devices connected to the network, such as the external printer, Ethernet hub and computer,
need to be grounded.
• Use a network cable with no damage to areas such as the covering.
• Do not place devices such as the printer, Ethernet hub or computer on the trolley.
• The network must be used only for this device and must be separate from other hospital networks.
• If two or more of this device are connected to the same network, a separate IP address must be assigned
to each device. (Set the IP address in the network settings of the device in "Network and Master ID
Settings".)
• Software such as VSS is required to connect the device to a computer.
WARNING
External Printer
Ethernet Hub
Main
Unit
Содержание VaSera VS-2000 Series
Страница 4: ......
Страница 22: ...18 ...
Страница 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Страница 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Страница 172: ...9 22 Chapter 9 File Transfers Communication History ...
Страница 176: ...10 4 Chapter 10 Recording Daily Reports ...
Страница 214: ...12 12 Chapter 12 Maintenance and Inspection ...
Страница 233: ...The company and product names used in this manual are trademarks or registered trademarks of respective companies ...