There is not enough space
on USB mem. Please
replace the media with
another one.
There is not enough space on the USB
memory to save the data.
Replace with a USB memory with enough capacity.
To not run out of battery,
the power supply will be
switched off automatically.
During battery operation, the power will
automatically turn OFF if the operation is
not performed within the specified duration.
Turn ON the power again.
Message
Cause
Countermeasure
Valve error
Cuff depressurization or deflation took too
long.
Inspect the cuff and air hose for damage or
breakage.
Message
Cause
Countermeasure