13-17
Chapter 13 Appendix
Item
Inspection Item
Procedure and
Criteria
Judgment
Remarks
(Repair
Needed)
Perfor-
mance
Blood
pressure
R-B
L-B
R-A
L-A
1
Zero point
Within 0±1mmHg
Pass/Fail
Pass/Fail Pass/Fail Pass/Fail
2
Pressure
accuracy
50
mmHg
Within ±3mmHg
Pass/Fail
Pass/Fail Pass/Fail Pass/Fail
100
mmHg
Within ±3mmHg
Pass/Fail
Pass/Fail Pass/Fail Pass/Fail
150
mmHg
Within ±3mmHg
Pass/Fail
Pass/Fail Pass/Fail Pass/Fail
250
mmHg
Within ±3mmHg
Pass/Fail
Pass/Fail Pass/Fail Pass/Fail
300
mmHg
Within ±3mmHg
Pass/Fail
Pass/Fail Pass/Fail Pass/Fail
3
Maximum pressure
320±10mmHg
Pass/Fail
Pass/Fail Pass/Fail Pass/Fail
4
Inflation time
20 sec. or less
Pass/Fail
Pass/Fail Pass/Fail Pass/Fail
5
Air leakage
6mmHg/min or less
Pass/Fail
Pass/Fail Pass/Fail Pass/Fail
6
Quick exhaust
10 sec. or less
Pass/Fail
Pass/Fail Pass/Fail Pass/Fail
7
100% open
100 to 250mmHg/sec. Pass/Fail
Pass/Fail Pass/Fail Pass/Fail
8
75% open
20 to 190mmHg/sec.
Pass/Fail
Pass/Fail Pass/Fail Pass/Fail
9
Time Limit
130 ±10sec., 8±1sec.
at high speed
Pass/Fail
Pass/Fail Pass/Fail Pass/Fail
BPU-
100
(Toes)
Right toe
Left toe
1
Zero point
Within 0±1mmHg
Pass/Fail
Pass/Fail
2
Pressure
accuracy
50
mmHg
Within ±3mmHg
Pass/Fail
Pass/Fail
100
mmHg
Within ±3mmHg
Pass/Fail
Pass/Fail
150
mmHg
Within ±3mmHg
Pass/Fail
Pass/Fail
250
mmHg
Within ±3mmHg
Pass/Fail
Pass/Fail
300
mmHg
Within ±3mmHg
Pass/Fail
Pass/Fail
3
Maximum pressure
Should be
320±10mmHg.
Pass/Fail
Pass/Fail
4
Inflation time
20 sec. or less
Pass/Fail
Pass/Fail
5
Air leakage
6mmHg/min or less
Pass/Fail
Pass/Fail
6
Quick exhaust
10 sec. or less
Pass/Fail
Pass/Fail
7
100% open
100 to 250mmHg/sec
Pass/Fail
Pass/Fail
8
75% open
20 to 190mmHg/sec
Pass/Fail
Pass/Fail
9
Time Limit
130 ±10sec., 8±1sec.
at high speed:
Pass/Fail
Pass/Fail
Item
Inspection Item
Procedure and Criteria
Judgment
Remarks
(Repair Needed)
Perfor-
mance
PCG
1
PCG filter charac-
teristics (PWV)
PCG filter
characteristics
165Hz
70% or above
Pass/Fail
280Hz
70% or above
Pass/Fail
2
PCG filter
L characteristics
30Hz
50% or above
Pass/Fail
50Hz
70% or above
Pass/Fail
80Hz
90% or above
Pass/Fail
Содержание VaSera VS-2000 Series
Страница 4: ......
Страница 22: ...18 ...
Страница 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Страница 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Страница 172: ...9 22 Chapter 9 File Transfers Communication History ...
Страница 176: ...10 4 Chapter 10 Recording Daily Reports ...
Страница 214: ...12 12 Chapter 12 Maintenance and Inspection ...
Страница 233: ...The company and product names used in this manual are trademarks or registered trademarks of respective companies ...