5-24
Chapter 5 Blood Pressure and Pulse Wave Examinations (CAVI/ABI/TBI): Standard Mode
Measuring CAVI (Manually)
During CAVI examinations, it is possible to obtain up to 100 seconds of waveform data and select and
measure any continuous heart rate from the screen while viewing the CAVI indicator.
This function is common to 2-channel, 4-channel and 6-channel examinations.
Below is the procedure for taking measurements.
When performing 2-channel examination on both sides, the measurement process needs to be performed
twice with separate measurements taken for the right and left side.
*The device returns to the standby status if it is paused for maximum of 100 seconds.
Manually Capturing and Measuring Waveforms
1
Touch the [Auto/Manual] button at the
top of the display to switch to manual
capture mode.
"Auto/Manual" is displayed at the top of the home
display when the device is in the waveform capture
mode.
2
Perform steps 3 to 9 of "Measuring
CAVI/ABI/TBI" (P5-9 ).
3
Press the [START] key on the display
unit.
4
"Press the START button to acquire the
waveforms" is displayed when
measurements can be taken.
5
Touch the [Freeze] button after the
boundaries of the waveform have been
displayed.
The acquired waveforms will be displayed. Scroll
through the displayed waveforms and select a
continuous waveform with no irregularities such as
arrhythmia by surrounding it with a white box. The
quality of the selected waveform will be displayed in
real time by the CAVI indicator.
The maximum waveform length that can be acquired
during freeze status is 100 seconds.
25
Press
[START].
CAVI Measurement
End of
waveform acquisition.
Press [START].
ABI/TBI Measurement
Start of
Measurement
End of
Measurement
Maximum 100 seconds.*
BP measurement
Touch [Freeze], and
select waveform.
Examination
Result Display
Save/Send
File
Color Report
Содержание VaSera VS-2000 Series
Страница 4: ......
Страница 22: ...18 ...
Страница 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Страница 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Страница 172: ...9 22 Chapter 9 File Transfers Communication History ...
Страница 176: ...10 4 Chapter 10 Recording Daily Reports ...
Страница 214: ...12 12 Chapter 12 Maintenance and Inspection ...
Страница 233: ...The company and product names used in this manual are trademarks or registered trademarks of respective companies ...