11-6
Chapter 11 Settings of This Device
Equipment Control Settings
The text in [ ] indicates the factory default setting.
General/Power
Setup Item
Description
Setting Range
Examination mode
Specifies the examination mode.
(Displayed only during 2-channel examination.)
The Simple Examination Mode is available only when English is selected as
the language.
[Standard Examination Mode],
Simple Examination Mode
Terminal Number
Specifies the terminal number of the device. This is useful when creating
system configurations.
[0]-255
Auto power-off
Turns the power off automatically if no operations are performed for the
prescribed period of time when using the battery.
OFF, [5 minutes], 10 minutes
Power Frequency
Specifies the frequency of the AC power supply to which the device is connected.
50Hz, [60Hz]
Password*
A password for entering the settings function (maximum of 4 digits) can be set. Not set