12-7
Chapter 12 Maintenance and Inspection
1. Safety specifications of this device
Classification by protection type: Class I equipment
Classification of applied part by degree of protection: Type CF applied part
2. Earth leakage current inspection
1) Measurement wiring diagram
2) Allowable values
3. Touch current inspection
1) Measurement wiring diagram
Exposed Metal: Potential equalization terminal, etc.
2) Allowable values
4. Patient leakage current (from patient connection to earth) inspection
1) Measurement wiring diagram
Patient Applied Part: Cuffs, PCG microphone chip, etc.
Normal condition
Single fault condition
5mA
10mA
Normal condition
Single fault condition
0.1mA
0.5mA
Medical Electrical
Equipment
AC
Outlet
AC
Outlet
Safety Evaluation
Test Device
(Leakage Current)
AC
Outlet
AC
Outlet
Medical Electrical
Equipment
Safety Evaluation
Test Device
(Leakage Current)
Exposed Metal Part
Contact Probe
AC
Outlet
AC
Outlet
Medical Electrical
Equipment
Contact Probe
Patient Applied Part
Safety Evaluation
Test Device
(Leakage Current)
Содержание VaSera VS-2000 Series
Страница 4: ......
Страница 22: ...18 ...
Страница 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Страница 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Страница 172: ...9 22 Chapter 9 File Transfers Communication History ...
Страница 176: ...10 4 Chapter 10 Recording Daily Reports ...
Страница 214: ...12 12 Chapter 12 Maintenance and Inspection ...
Страница 233: ...The company and product names used in this manual are trademarks or registered trademarks of respective companies ...