12-8
Chapter 12 Maintenance and Inspection
2) Allowable values
5. Patient leakage current (external voltage) inspection
1) Measurement wiring diagram
Patient Applied Part: Cuffs, PCG microphone chip, etc.
2) Allowable values
6. Patient auxiliary current inspection
1) Measurement wiring diagram
Patient Applied Part: Cuffs, PCG microphone chip, etc.
2) Allowable values
Normal condition
Single fault condition
Direct Current
0.01mA
0.05mA
Alternating Current
0.01mA
0.05mA
Single fault condition
0.05mA
Normal condition
Single fault condition
Direct Current
0.01mA
0.05mA
Alternating Current
0.01mA
0.05mA
AC
Outlet
AC
Outlet
Medical Electrical
Equipment
Patient Applied Part
Contact Probe
Safety Evaluation
Test Device
(Leakage Current)
AC
Outlet
AC
Outlet
Medical Electrical
Equipment
Patient Applied Part
Contact Probe
Safety Evaluation
Test Device
(Leakage Current)