Precautions about Transportation
C A U T I O N
When transporting this equipment, pack it with specified packing materials.
Also, transport it under appropriate environment condition.
Monitoring after Power Failure
When the power failure is within 30 seconds, monitoring will resume with the display mode and patient
information unchanged. When the power failure is 30 seconds or more, monitoring will resume with the
default display mode of factory setting or user setting, or the display mode which was last set, only if the
equipment was operated for 30 minutes or more before the power failure.
HPD-800/HPD-810, HCP-800/HCP-810 will start up from the warm-up mode. The warm-up time differs
for each unit.
To Prepare for Emergency Use
Accessories/Optional Accessories
z
The ECG electrodes are consumable products. Always prepare extra supplies of electrodes.
z
Check once a week that there is no wire break on the patient cable.
Battery Pack
z
Even if the battery pack is not in use, the remaining capacity decreases due to self-discharge. Make
sure to verify periodically that the battery pack is fully charged
z
To fully charge the empty battery pack, it takes 8 hours during operation, and 3.5 hours when the
power is OFF and AC cable is connected.
z
The performance of the battery deteriorates with repeated use. To ensure performance of the battery,
it is recommended to replace it once a year.
xxvii
Содержание DS-8200
Страница 4: ...Blank Page ...
Страница 44: ...1 4 Blank Page ...
Страница 112: ...4 40 Blank Page ...
Страница 113: ...Wiring Diagram This chapter describes the wiring diagram of the equipment Chapter 5 DS 8200 System 5 2 5 1 ...
Страница 116: ...5 4 Blank Page ...
Страница 156: ...8 4 Blank Page ...
Страница 166: ...9 10 Blank Page ...
Страница 208: ...11 4 ...