DIP Switch/Rotary Switch Setup
This section describes the settings of the DIP switch and rotary switch.
DIP Switch Setup
The DIP switch is located inside the maintenance cover of the LC-8210.
SW No.
Default
Description (When ON)
1
OFF
Not used. (Reserved function)
2
OFF
Not used. (Reserved function)
3
OFF
Not used. (Reserved function)
4
OFF
Not used. (Reserved function)
5 ON
SpO
2
body motion transmission during telemetry transmission
Not transmit NIBP invalid value.
6
OFF
Release the hardware write protect of the flash memory.
7
OFF
Tone when the power is cut OFF
8 OFF
Forced application software installation and
EEPROM data writing
5: Sets ON/OFF of SpO
2
body motion transmission during telemetry transmission.
The setup depends on the telemetry center capability of SpO
2
body motion information.
6: Set up only when installing the software (upgrading boot program).
Prevents unintended installation. Default setting is OFF.
7: When the AC power cable is disconnected or the power supply SW on the rear side of the
main unit is OFF, a beep sound will be generated. The sound notifies unintended
interruption of the power supply, and the default setting is OFF.
8: Uses only for installing the software/EEPROM.
It prevents unintended installation. Default setting is OFF.
10-2
Содержание DS-8200
Страница 4: ...Blank Page ...
Страница 44: ...1 4 Blank Page ...
Страница 112: ...4 40 Blank Page ...
Страница 113: ...Wiring Diagram This chapter describes the wiring diagram of the equipment Chapter 5 DS 8200 System 5 2 5 1 ...
Страница 116: ...5 4 Blank Page ...
Страница 156: ...8 4 Blank Page ...
Страница 166: ...9 10 Blank Page ...
Страница 208: ...11 4 ...