C A U T I O N
z
z
When the patient's data may become mixed with a different patient's data due to
interference.
y
When there are multiple TCON communication devices set to the same TCON
ID and channel (group).
z
When symptoms such as being unable to communicate, unstable
communication, or poor reception may occur.
y
When the radio communication is bad because there are metal, concrete, or
other such obstacles between the Bidirectional Wireless Communications
Modules (TCON).
y
When a different wireless device is using the same frequency (channel).
y
When there are other TCON devices nearby using different channels (groups).
y
When a cell telephone or other wireless device is being used nearby.
y
When citizens broadcast bands such as amateur radio or truck radios are used
in the vicinity of the TCON operating area.
y
When a computer or word processor, or electrical device that has an internal
computer, is used near the TCON device antenna.
y
When the TCON device is installed or moved to a location that is outside the
radio communication range.
y
If a nearby different group is set with a TCON channel frequency that is too
close to the channel frequency set for the current TCON group.
Precautions about the Setting
Follow the instructions from the Overall Manager for the wireless channel when
setting the TCON and channel IDs to prevent interference within the same
institution.
z
If the TCON is set to [OFF], all TCON messages such as "Check TCON Comm."
will not be displayed.
z
When TCON is used, even if [ON] is set for "Start NIBP Auto Mode with
Start/Stop key."([Initial Settings]>[User I/F]>[Power ON/Discharge]), Backup at
Discharge (NIBP Auto Mode) function will be [ON] since the central monitor will
not be on Standby mode.
z
Check that three antenna bar marks (
) are displayed.
z
Make sure that the TCON group number between the bedside monitor and
central monitor is the same.
z
If the equipment is moved during TCON operation, the radio waves signal may
become interrupted.
z
There are following restrictions when connecting the DS-8200 System to the
TCON network.
y
When the BP measurement unit is [kPa], the central monitor does not receive
the NIBP/BP1/BP2 Data. Also, the NIBP/BP1/BP2 Alarm Setup cannot be
changed on the central monitor.
y
The NIBP measurement cannot be started from the central monitor via TCON
system if the NIBP measurement interval is set to [5 min] or less, or during the
1-minute or continuous measurement. However, it can be stopped.
y
If the measurement unit of CO
2
concentration is [mmHg], the CO
2
value of
100mmHg or above will be transmitted as 99mmHg.
xii
Содержание DS-8200
Страница 4: ...Blank Page ...
Страница 44: ...1 4 Blank Page ...
Страница 112: ...4 40 Blank Page ...
Страница 113: ...Wiring Diagram This chapter describes the wiring diagram of the equipment Chapter 5 DS 8200 System 5 2 5 1 ...
Страница 116: ...5 4 Blank Page ...
Страница 156: ...8 4 Blank Page ...
Страница 166: ...9 10 Blank Page ...
Страница 208: ...11 4 ...