C A U T I O N
z
Precautions for Single-Patient-Use Type Sensors
The sensor can be reused on the same patient as long as the adhesive tape
attaches without slippage. But do not reuse on other patients to avoid cross
contamination. It is intended for single patient use only. For additional warnings,
cautions, or contraindications when using sensors with Nellcor Unit
(HS-8312N)/Masimo Unit (HS-8312M), refer to SpO
2
sensor instruction manual.
z
If "- - -" is displayed for the SpO
2
numeric data, make sure that the sensor is
properly attached.
z
Measuring on a limb with NIBP cuff, arterial catheter, or intracatheter may
result in incorrect measurement.
z
Venous congestion may cause under reading of actual oxygen saturation.
Therefore, assure proper venous outflow from monitored site. Sensor should
not be below heart level (e.g. sensor on hand of a patient in a bed with arm
dangling to the floor).
Precautions about the NIBP Monitoring
z
Pay attention when measuring the NIBP of patient with bleeding disorders or
hyper coagulation. The cuff inflation may cause petechia or circulatory failure
by the blood clot.
z
For the following situation, measurements will be terminated.
y
When the measurement time has exceeded 160 seconds for adult and child,
80 seconds for neonate.
y
When the inflation value has exceeded 300mmHg for adult, 210mmHg for
child, and 150mmHg for neonate.
z
If used with the incorrect patient classification, it will not only cause erroneous
measurement, but the inflating level for the adult may be applied to child or
neonate causing dangerous situation to the patient.
z
The continuous measurement and 1-minute interval measurement will
automatically stop after 12 minutes (maximum 15 minutes).
z
If the mean MAP display is set to OFF, the MAP alarm will not be generated.
Also the MAP data will not be displayed for the tabular trend or the NIBP list.
Precautions about the BP Monitoring
z
If the transducer gets disconnected, pay attention that the metal part of the
transducer does not get in touch with any metal parts of the bed or any
conductive parts. Also, the operator should not touch the conductive parts with
bare hands. Otherwise, it may result in electric shock to the patient and/or
operator.
z
When the power is turned ON, the BP value will not be displayed until zero
balance is performed. Make sure to perform the zero balance. Once the zero
balance is performed, the zero balance information will be maintained, and the
BP value will be displayed.
z
Each time the blood pressure transducer or tubing is replaced, the zero
balance procedure is required to ensure accurate measurements.
z
The zero balance procedure is required for the following case.
y
When starting the measurement.
y
When the position of the heart has changed due to body movement.
y
When the position of the transducer has changed.
y
When measuring for a long period of time and there is a possibility of
measurement error due to change in ambient temperature, etc.
y
When a connector is connected/disconnected, or a transducer is replaced.
z
Note that Systolic Pressure (SYS) = Peak Systolic Pressure (PSP) for the
graphic trend, data base, and alarm setup.
z
When ECG is not measured, PDP cannot be calculated.
z
The undisplayed BP data (SYS/DIA/Mean) will not generate a BP alarm or be
displayed in the tabular trend. Select the appropriate display type according to
the monitoring purpose.
xxi
Содержание DS-8200
Страница 4: ...Blank Page ...
Страница 44: ...1 4 Blank Page ...
Страница 112: ...4 40 Blank Page ...
Страница 113: ...Wiring Diagram This chapter describes the wiring diagram of the equipment Chapter 5 DS 8200 System 5 2 5 1 ...
Страница 116: ...5 4 Blank Page ...
Страница 156: ...8 4 Blank Page ...
Страница 166: ...9 10 Blank Page ...
Страница 208: ...11 4 ...