C A U T I O N
Precautions about the System Setup
z
When the waveform and numeric data display for each parameter is set to
OFF, the alarm and trend input will be also suspended.
z
If the HR/PR source is set to [BP], and if BP waveform/numeric data is set to
[Disp. OFF], the PR value will not be displayed.
z
If the HR/PR source is set to [SpO
2
], and if SpO
2
waveform/numeric data is set
to [Disp. OFF], the PR value will not be displayed.
z
If the RR source is set to [CO
2
/GAS], and if CO
2
waveform/numeric data is set
to [Disp. OFF], the RR value will not be displayed.
z
Do not set the same remote control bed ID to more than one monitors on the
same floor. Other
z
After the remote control setup, check that the remote control unit is properly
operating.
z
If the time/date is not correctly set, or changed during monitoring, malfunction
may occur with NIBP measurement, periodic recording, trend, NIBP list data,
and age calculation from the birth date.
Precautions about the Patient Admit/Discharge
z
If monitoring of a new patient is started without discharging the previous
patient, data of the new patient will be added to the data of the previous patient
which will result in inaccuracy.
z
The user mode setting (alarm/display configuration) will remain effective even
when the power is turned OFF or when the patient is discharged. Before
monitoring, make sure the current user mode is suitable for the patient's
condition.
z
Resuming monitoring will also resume the alarm in suspension.
Precautions about the CF/SD Card
z
Use only the specified CF/SD card.
z
Use only the CF/SD card formatted with this equipment.
z
Make sure to turn the power OFF and ON again after the setup data is read
from the CF/SD card.
The read setup data will become effective after the power is turned OFF and
ON again.
z
Reading the patient data from the CF/SD card will erase all previous patient
data stored in the patient monitor.
Precautions about the Maintenance
z
When cleaning the touch panel, never use strong-acidic cleaning solution.
z
A special coating is applied to the surface of the touch panel. Do not wipe the
surface with a cloth or gauze with coarse texture. Wipe the surface with the soft
cleaning cloth provided as optional accessory or with an eyeglass cleaning
cloth.
z
Clean the equipment frequently so stains can be removed easily.
z
To prevent injury, it is recommended to wear gloves when cleaning the
equipment.
z
Do not allow liquids or cleaning solution to enter the equipment or connectors.
z
Do not use organic solvents, thinner, toluene and benzene to avoid damaging
the resin case.
z
Do not polish the housing with abrasive or chemical cleaner.
z
When sterilizing the entire room using a spray solution, pay close attention not
to have liquids get into the equipment or connectors.
z
The surface resin coating may be damaged, resulting in discoloration,
scratches, or other problems.
Example:
Do not use chemical cloth, scrub brush, abrasive, polishing powder, hot water,
volatile solvent and chemicals (cleanser, thinner, benzine, benzol, and synthetic
detergent for house and furniture), or sharp-edged tools.
z
Do not open the housing.
z
Avoid alcohol or other liquids from getting into the equipment.
z
Replace the periodic replacement parts periodically as specified.
xxiii
Содержание DS-8200
Страница 4: ...Blank Page ...
Страница 44: ...1 4 Blank Page ...
Страница 112: ...4 40 Blank Page ...
Страница 113: ...Wiring Diagram This chapter describes the wiring diagram of the equipment Chapter 5 DS 8200 System 5 2 5 1 ...
Страница 116: ...5 4 Blank Page ...
Страница 156: ...8 4 Blank Page ...
Страница 166: ...9 10 Blank Page ...
Страница 208: ...11 4 ...