Precautions when Using with Other Equipment
Pacemaker
W A R N I N G
z
Minute ventilation rate-adaptive implantable pacemakers can occasionally
interact with certain cardiac monitoring and diagnostic equipment, causing the
pacemakers to pace at their maximum programmed rate. The cardiac
monitoring and diagnostic equipment may possibly send wrong information. If
such event occurs, please disconnect the cardiac monitoring and diagnostic
equipment, or follow the procedures described in the operation manual of the
pacemaker. For more details, contact FUKUDA DENSHI personnel, your
institution's professionals, or your pacemaker distributors.
z
Rate meters may continue to count the pacemaker rate during occurrences of
cardiac arrest or some arrhythmias. Do not rely entirely upon rate meter
alarms. Keep pacemaker patients under close surveillance.
Reference
“Minute Ventilation Rate-Adaptive Pacemakers”
FDA alerts health professionals that minute ventilation rate-adaptive
implantable pacemakers can occasionally interact with certain cardiac
monitoring and diagnostic equipment, causing pacemakers to pace at
their maximum programmed rate.
[Based on a safety bulletin issued by FDA Center for Devices and
Radiological Health on October 14, 1998]
Non-Explosion Proof
D A N G E R
Never operate the equipment in the presence of flammable anesthetics, high
concentration of oxygen, or inside hyperbaric chamber. Also, do not operate the
equipment in an environment in which there is a risk of explosion.
Explosion or fire may result.
Defibrillator
W A R N I N G
z
When defibrillating, keep away from the electrodes or medicament applied to
the patient chest. If this is not possible, remove the electrodes or medicament
before defibrillating.
If the defibrillator paddles are directly in contact with the electrodes or
medicament, an electrical shock may result by the discharged energy.
z
When defibrillating, make sure that the electrodes, sensor cables, or relay
cables are firmly connected to the device.
Contacting the metal part of the disconnected cable may result in electrical
shock from the discharged energy.
z
When defibrillating, do not touch the patient and the metal part of the device or
cables.
Electric shock may result from the discharged energy.
z
This equipment will return to standard operating mode within 10 seconds after
defibrillating. The stored data will not be affected. The measurement accuracy
will temporarily decrease during defibrillation, but it will not compromise the
safety of patient and the equipment.
z
The QRS synchronized signal is not intended to be used as synchronized
signal for defibrillator.
xiii
Содержание DS-8200
Страница 4: ...Blank Page ...
Страница 44: ...1 4 Blank Page ...
Страница 112: ...4 40 Blank Page ...
Страница 113: ...Wiring Diagram This chapter describes the wiring diagram of the equipment Chapter 5 DS 8200 System 5 2 5 1 ...
Страница 116: ...5 4 Blank Page ...
Страница 156: ...8 4 Blank Page ...
Страница 166: ...9 10 Blank Page ...
Страница 208: ...11 4 ...