Base Unit: BS-8210
Size
270(W)×180(D)×92(H)mm (not including the protrusion)
Weight
2.5kg (not including the accessory)
Environmental Conditions
Operating Temperature : 10 to 40°C
Operating Humidity
: 30 to 85 % (non-condensing)
Transport / Storage
: -10 to 60°C
Temperature
Transport / Storage
: 10 to 95% (40°C) (non-condensing)
Humidity
However, for the CF-820 IR Remote Control Unit, the following
condition applies.
10 to 90% (38°C) (non-condensing)
Storage Atmospheric
: 700 to 1060hPa
Pressure
Vibration Proof
: Comply with MIL-STD-810G:2008 METHOD514.6 VIBRATION
Safety
General Standard
: EN 60601-1 : 1990+A1: 1993+A2: 1995 (for Europe)
IEC 60601-1:1988+A1:1991+A2:1995 (for USA)
(Medical Electrical Equipment- Part 1: General Requirements for Safety)
EN 60601-1-1: 2001 (for Europe)
IEC 60601-1-1:2000 (for USA)
(Medical Electrical Equipment- Part 1-1: General Requirements for Safety
-Collateral standard: Safety Requirements for Medical Electrical Systems)
EMC Standard
: EN 60601-1-2 (2012) (for Europe)
IEC 60601-1-2: 2007 (for USA)
(Medical Electrical Equipment- Part 1-2: General Requirements for Safety
–Electromagnetic
-
Compatibility Requirements and Tests)
Type of protection against: Class I Equipment (During AC power operation)
electric shock
Internally powered Equipment (During battery operation)
Operation Mode
:
Continuous Operating Equipment
The degree of protection
:
Combination of LC-8210, HSB-80, HS-8000 and BS-8210: IPX1
against ingress of water
Other situation: IPX0
Protection against Ignition: Not provided
of Frammable Gas
Power Supply (During AC power operation)
Voltage : AC 100 to 240V
Frequency : 50/60 Hz
Power Consumption : 80VA
Power Supply (During battery operation)
Voltage
: DC14.8V (Lithium-ion Battery Pack BTO-008)
Power Consumption
: When LC-8210, HSB-80, HS-8000 and BS-8210 are connected: 70W
Battery Operation Time : One battery (HSB-80): 2.5 hours (during measurement of NIBP 15min.
interval, ECG, SpO
2
)
Two batteries (HSB-80 and BS-8210): 5 hours (during measurement of
NIBP 15min. interval, ECG, SpO
2
)
The battery of BS-8210 will be used in priority.
Battery Charging Time
: Rapid Charge (when the equipment is not operating): 3.5 hours
Normal Charge (when the equipment is operating): 8 hours
During AC power operation, batteries of HSB-80 and BS-8210 will be
charged
simultaneously.
Usable Life
6 years
: According to self-certification.
2-4
Содержание DS-8200
Страница 4: ...Blank Page ...
Страница 44: ...1 4 Blank Page ...
Страница 112: ...4 40 Blank Page ...
Страница 113: ...Wiring Diagram This chapter describes the wiring diagram of the equipment Chapter 5 DS 8200 System 5 2 5 1 ...
Страница 116: ...5 4 Blank Page ...
Страница 156: ...8 4 Blank Page ...
Страница 166: ...9 10 Blank Page ...
Страница 208: ...11 4 ...