Test Menu Items (Cont.)
Description
Log 1 Log 2 Log 3
Records the time and information at power ON/OFF, abnormal
system restart, exceptional handling. This information can be
used to find solution when equipment failure occurs.
Log 1 (HS-8000)
Records the time and information at power ON/OFF, abnormal
system restart, exceptional handling. This information can be
used to find solution when equipment failure occurs.
Printer Adjustment
When the printer unit is replaced or repaired, readjustment will
be performed. Also, fine adjustment of the print density can be
performed. The adjustment result will be stored. For details,
refer to the section on “Printer Adjustment”.
NIBP Measurement
This is the test function for development.
NIBP Test
This is the test mode for the NIBP pump. A service tools is
required to perform this test. For details, refer to “Chapter 9
Maintenance” of “Super Unit HS-8000 Series Service Manual”.
* On this test mode, never press the [START] key when
the cuff is connected to the patient.
Arrhythmia Analysis
Displays the sample data for arrhythmia analysis and the data
of the analyzed result.
QRS Detect
If the QRS detection is not performed correctly with double
trigger signals, record each waveform for Mode 1, Mode 2,
and Mode 3, and then contact Fukuda Denshi.
Analog Parameter
Displays the numeric data during parameter measurement.
PERFORMANCE
Interrupt data will be displayed in increment.
This information can be used to find solution when equipment
failure occurs.
Seamless Log
Displays the event log related to the seamless function.
This information can be used to find solution when equipment
failure occurs.
Software Dip-SW
Sets up each software and Dip-SW.
WDT Test
WDT test is used for factory inspection. Normally, do not press
this key. All the patient data that has been set up will be
cleared.
HS-8000 WDT Test
WDT test is used for factory inspection. Normally, do not press
this key. All the patient data that has been set up will be
cleared.
Demo Start WDT Test
This key will start the DEMO mode.
The DEMO mode key will be displayed only when the rotary
switch on the main body is set to "2" or "7".
When the demonstration is finished, please make sure to turn
off the power and return the rotary switch to "0". All the patient
data that has been set up will be cleared.
This key will not function when the ECG cable is connected.
Analogue Adjust (Default)
This adjusts the analog board.
A special tool is required to adjust the analog board.
For details, refer to “Chapter 9 Maintenance” of “Super Unit
HS-8000 Series Service Manual”.
Initialize Password
All the passwords that have been set up will be cleared.
Initialize
All the data that has been set up will be cleared.
10-6
Содержание DS-8200
Страница 4: ...Blank Page ...
Страница 44: ...1 4 Blank Page ...
Страница 112: ...4 40 Blank Page ...
Страница 113: ...Wiring Diagram This chapter describes the wiring diagram of the equipment Chapter 5 DS 8200 System 5 2 5 1 ...
Страница 116: ...5 4 Blank Page ...
Страница 156: ...8 4 Blank Page ...
Страница 166: ...9 10 Blank Page ...
Страница 208: ...11 4 ...