C A U T I O N
z
On the central monitor, the respiration waveform and RR value based on the
RR/APNEA alarm source selected on the DS-8200 System will be displayed.
The RR and APNEA monitored on the central monitor and the DS-8200 System
will be the same.
Wireless Network System
D A N G E R
When monitoring a patient using wireless telemetry, make sure the patient data is
properly received at the central monitor. Pay special attention when the channel
ID at the bedside monitor is changed.
W A R N I N G
z
A password can be set to access the channel ID setup menu to allow only the
telemetry channel administrator to change the channel ID.
z
Some wireless combinations of telemetry transmitters may generate
interference with other devices.
z
Before selecting the channel, verify it will not interfere with other channels.
z
Make sure the telemetry manager of your system is aware of any changes to
the telemetry channels.
z
If transmitters are used in a neighboring medical facility, your facility and the
neighboring facility must make agreements on the setting of the telemetry
channels to prevent telemetry interference.
C A U T I O N
Precautions about the Telemetry
z
When performing telemetry transmission, configure the display so that the
numeric data corresponded to the waveform is displayed. If not, the displayed
waveform or numeric data may not be transmitted.
z
The setup of channel ID and group ID should be performed only by the
telemetry channel administrator or our service representative. Users should not
perform this procedure as malfunction may occur.
z
When the measurement unit of BP is "kPa", corresponding waveform and
numeric data will not be transmitted. When using a wireless network, use
"mmHg" for the BP measurement unit.
z
BP waveform with a scale above the set scale can not be properly transmitted.
When transmitting the BP waveform, pay attention to the displayed BP scale.
z
If the measurement unit of CO
2
concentration is mmHg and [99mmHg] is
selected for "Upper Limit of CO
2
(mmHg) Transmission" ([Initial
Settings]>[System]>[Telemeter]), the CO
2
value of 100mmHg or above will be
transmitted as 99mmHg.
RTC and Data Backup
C A U T I O N
z
This equipment is equipped with a built-in clock. When the power of this
equipment is turned OFF, this clock is backed up by a lithium primary battery. If
incorrect time is displayed when turning ON the power, a low battery may be
the cause. In such case, contact Fukuda Denshi for replacing the battery.
z
To protect the data during voltage dip, short interruptions and voltage
variations on power supply input lines or during short duration of power turned
OFF, this equipment performs 5-minute (approx.) data backup using the
secondary battery. The data may not be protected if the power is turned OFF
within 30 minutes from power ON.
xxv
Содержание DS-8200
Страница 4: ...Blank Page ...
Страница 44: ...1 4 Blank Page ...
Страница 112: ...4 40 Blank Page ...
Страница 113: ...Wiring Diagram This chapter describes the wiring diagram of the equipment Chapter 5 DS 8200 System 5 2 5 1 ...
Страница 116: ...5 4 Blank Page ...
Страница 156: ...8 4 Blank Page ...
Страница 166: ...9 10 Blank Page ...
Страница 208: ...11 4 ...