How to Use
WARNING
• Use an electrosurgical unit conforming to EN 60601-2-2 (IEC 60601-2-2). If any other
electrosurgical unit is used, it may cause severe harm to patient and/or end-users.
• Use the electrosurgical unit and related endotherapy devices per instructions provided in
the operation manual for each respective device. Otherwise, device failure, electric shock
and/or burns may occur.
• Prior to performing endoscopic electrosurgery, one should have a thorough understanding
of the manufacturers' instructions for all equipment involved and one should be familiar
with the specific safety and usage aspects of each endotherapy device.
CAUTION
• Prevent patient’s body from touching electric conductor such as metal part of bed while
performing high-frequency treatment. It could cause thermal injury to a patient due to
energization via the conductive part.
• When performing high-frequency treatment, take care that patient’s vomitus or body fluids
do not make contact with the conductive parts such as a metal part of the bed. It could
cause thermal injury to a patient due to energization via vomitus or body fluids.
• While performing high-frequency treatment, ensure that the end-user does not touch the
patient. It could cause thermal injury to a patient and/or end-user.
• Operate the electrosurgical unit within specified output range as per instructions provided
in the operation manual of the electrosurgical unit. Leakage current may cause thermal
injury.
• Do not energize the electrosurgical unit when the high-frequency endotherapy device or
electrically active portion is in contact with the distal end of endoscope. Thermal injury to a
patient or endoscope failure may occur.
• Do not apply the current under the circumstance that patient’s clothing is wet when
performing high-frequency treatment. Doing so may cause thermal injury.
(1)
Prepare, inspect and connect the electrosurgical unit and high-frequency endotherapy device
as per instructions provided in respective manuals.
(2)
Perform high-frequency treatment as per instructions provided in the manual of the
electrosurgical unit and high-frequency endotherapy device.
ED580XT_E2-30_897N200877B.indb 17
2018/09/20 15:50:25
Содержание ED-580XT
Страница 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Страница 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Страница 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Страница 35: ...897N200877B 2 3 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 3 2018 09 20 15 50 17...
Страница 48: ...897N200877B 2 16 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 16 2018 09 20 15 50 19...
Страница 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Страница 82: ...897N200877B 4 30 Chapter 4 Preparation and Inspection ED580XT_E2 30_897N200877B indb 30 2018 09 20 15 50 23...
Страница 114: ...897N200877B 6 12 Chapter 6 Troubleshooting ED580XT_E2 30_897N200877B indb 12 2018 09 20 15 50 25...
Страница 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Страница 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...