897N200877B
Appendix-9
Appendix
<Portable and Mobile RF Communications Equipment Compliance Information and Guidance>
Table 3
Guidance and Manufacturer Declaration - Electromagnetic Immunity -
This device is intended for use in the following prescribed electromagnetic environments. Customers and users of
this product are advised to check that it is being used in such environments.
Immunity test
IEC 60601
Test level
Compliance level
Guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
6 Vrms
ISM Frequency Band
c
3 Vrms
6 Vrms
ISM Frequency Band
c
Portable and mobile RF
communications equipment should
be used no closer to any part of this
product, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1.2
√
P
Radiated RF
IEC/EN 61000-4-3
3 V/m
80 MHz to 2.7 GHz
3 V/m
d = 1.2
√
P
80 MHz to 800 MHz
d = 2.3
√
P
800 MHz to 2.7 GHz
Where “P” is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and “d”
is the recommended separation
distance in meters (m). The
electric field intensity from a fixed
RF transmitter determined by an
electromagnetic field study
a
should
be lower than the compliance level
in each frequency range
b
.
Interference may occur near devices
that display the following symbol.
Immunity to
proximity fields
from RF wireless
communications
equipment
IEC/EN 61000-4-3
380 - 390 MHz, 27 V/m
430 - 470 MHz, 28 V/m
704 - 787 MHz, 9 V/m
800 - 960 MHz, 28 V/m
1422 - 1512 MHz, 10 V/m
1700 - 1990 MHz, 28 V/m
2400 - 2570 MHz, 28 V/m
3480 - 3600 MHz, 10 V/m
5100 - 5800 MHz, 9 V/m
380 - 390 MHz, 27 V/m
430 - 470 MHz, 28 V/m
704 - 787 MHz, 9 V/m
800 - 960 MHz, 28 V/m
1422 - 1512 MHz, 10 V/m
1700 - 1990 MHz, 28 V/m
2400 - 2570 MHz, 28 V/m
3480 - 3600 MHz, 10 V/m
5100 - 5800 MHz, 9 V/m
Degradation of the performance of
this product could result if portable
RF communications equipment is
used closer than 30 cm to any part
of this product.
Note
• At 80 MHz and 800 MHz, the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
ED580XT_E2-30_897N200877B.indb 9
2018/09/20 15:50:26
Содержание ED-580XT
Страница 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Страница 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Страница 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Страница 35: ...897N200877B 2 3 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 3 2018 09 20 15 50 17...
Страница 48: ...897N200877B 2 16 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 16 2018 09 20 15 50 19...
Страница 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Страница 82: ...897N200877B 4 30 Chapter 4 Preparation and Inspection ED580XT_E2 30_897N200877B indb 30 2018 09 20 15 50 23...
Страница 114: ...897N200877B 6 12 Chapter 6 Troubleshooting ED580XT_E2 30_897N200877B indb 12 2018 09 20 15 50 25...
Страница 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Страница 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...