CAUTION
• If resistance is encountered while advancing an endotherapy device within the instrument
channel, do not forcibly advance the endotherapy device. Otherwise, it may cause
malfunction of the endoscope.
• When inserting an endotherapy device, close the lid of the forceps valve. If the lid is open,
it can reduce the efficacy of the endoscope’s suction system, making it impossible to
perform aspiration.
Note
• Occasionally, an endotherapy device can encounter difficulty while attempting to
pass through an angulated bending section. In such case, reduce the angulation in
the bending section a little and try to advance the endotherapy device again.
• If the handle of an endotherapy device is held tightly, the endotherapy device may
not be inserted smoothly or the maximum bending angle may decrease. If this
happens, decrease the holding force.
• If an accessory cannot be advanced or meets resistance at the scope distal tip, make
sure that the forceps elevator is lowered so as not to impede or restrict movement of
the accessory.
For information on handling an endotherapy device, refer to the manual of the endotherapy
device. Use an endotherapy device given in this manual.
“Appendix - Related Equipment Used in Combination”
Note
• Do not open the distal end of an endotherapy device with a distal opening/closing
function while it is being inserted into the endoscope.
• Do not insert an endotherapy device with a distal needle into the endoscope while
the needle is protruding from the distal end.
• Do not protrude the needle of an endotherapy device with a distal needle while it is
being inserted into the endoscope.
• Do not detach the stopper of a clip before or while a detainment endotherapy device
is being inserted into the endoscope.
• Do not open the clip while a detainment endotherapy device is being inserted into
the endoscope.
ED580XT_E2-30_897N200877B.indb 15
2018/09/20 15:50:25
Содержание ED-580XT
Страница 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Страница 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Страница 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Страница 35: ...897N200877B 2 3 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 3 2018 09 20 15 50 17...
Страница 48: ...897N200877B 2 16 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 16 2018 09 20 15 50 19...
Страница 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Страница 82: ...897N200877B 4 30 Chapter 4 Preparation and Inspection ED580XT_E2 30_897N200877B indb 30 2018 09 20 15 50 23...
Страница 114: ...897N200877B 6 12 Chapter 6 Troubleshooting ED580XT_E2 30_897N200877B indb 12 2018 09 20 15 50 25...
Страница 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Страница 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...