5.2 Insertion and Observation
5.2.1 Insertion
WARNING
• Do not forcibly advance or withdraw the endoscope into/from the patient, angulate the
bending section forcibly or operate it quickly. It may cause patient injury, bleeding or
perforation.
• If a patient sneezes or moves abruptly during the procedure, malfunction of the endoscope
and patient bleeding or trauma may occur. Depending on the degree of malfunction, safe
endoscope withdrawal may be difficult or impossible, causing severe harm to patient and/
or end-users.
CAUTION
• Do not directly apply Xylocaine spray to the insertion portion. Do not use olive oil as a
lubricant for insertion. It may cause deterioration of the outer surface.
• Securely attach the distal end cap to the distal end of the endoscope before use. If the
distal end cap is not attached securely, the distal end cap may fall off during use.
• Use this product in combination with the distal end cap DC-07D. If another type of distal
end cap is used in combination with this product, the distal end cap may fall off during use.
• If the distal end cap becomes misaligned or falls off during a procedure, stop the
procedure immediately and slowly withdraw the endoscope. Otherwise, it may cause injury
to tissues in the body cavity.
• If you encounter any resistance during a procedure, insert the endoscope slowly. Do not
force it in. Do not insert or bend the endoscope without securing the view on the monitor.
Not following the recommendations above may cause endoscope failure or patient injury.
• Lower the forceps elevator before inserting the endoscope. Not doing so may cause
patient injury.
(1)
Patients should lie on examining table in an appropriate position as required for the specific
type of endoscopy procedure to be performed.
ED580XT_E2-30_897N200877B.indb 4
2018/09/20 15:50:24
Содержание ED-580XT
Страница 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Страница 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Страница 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Страница 35: ...897N200877B 2 3 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 3 2018 09 20 15 50 17...
Страница 48: ...897N200877B 2 16 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 16 2018 09 20 15 50 19...
Страница 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Страница 82: ...897N200877B 4 30 Chapter 4 Preparation and Inspection ED580XT_E2 30_897N200877B indb 30 2018 09 20 15 50 23...
Страница 114: ...897N200877B 6 12 Chapter 6 Troubleshooting ED580XT_E2 30_897N200877B indb 12 2018 09 20 15 50 25...
Страница 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Страница 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...