4.4 Inspecting and Attaching Accessories
Prior to inspection, make sure that dirt does not adhere to the accessories. If dirt adheres
to the accessories, reprocess them according to the instructions given in the Reprocessing
Manual.
4.4.1 Forceps Valve
<When the lid is open>
<When the lid is closed (top view)>
<When the lid is closed (side view)>
Slit
Slit
Circular Hole
Main Body of Forceps Valve
Lid
Main Body of
Forceps Valve
WARNING
• The forceps valve is intended for single use. Discard it after use. If a deteriorated forceps
valve is used, body fluids may leak, causing infection.
• Reprocess the forceps valve before use. Use of an improperly reprocessed forceps valve
can create a risk of infection.
• Make sure to check the forceps valve before use. If the inspection result shows any sign of
abnormality or irregularity, replace the forceps valve with a new one already reprocessed.
Use of an abnormal forceps valve may cause the leakage of body fluid, posing an infection
risk.
• Ensure that the forceps valve is properly attached to the instrument channel inlet. If this
product is used without the forceps valve attached, body fluid may leak and it could be a
source of infection.
<Preparation>
The forceps valve is not reprocessed before shipping from FUJIFILM. Reprocess it according
to the instructions given in the Reprocessing Manual before using it in a procedure.
Reprocessing Manual
<Inspection>
(1)
Confirm that the slit and circular hole on the forceps valve are free from abnormalities such as
splits, cracks, deformations or discoloration.
(2)
Close the lid and visually check that there is no clearance between the lid and the main body
of the forceps valve.
Forceps Valve
Suction Valve Air/Water Valve Distal End Cap
4.4 Inspecting and Attaching Accessories
ED580XT_E2-30_897N200877B.indb 13
2018/09/20 15:50:21
Содержание ED-580XT
Страница 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Страница 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Страница 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Страница 35: ...897N200877B 2 3 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 3 2018 09 20 15 50 17...
Страница 48: ...897N200877B 2 16 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 16 2018 09 20 15 50 19...
Страница 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Страница 82: ...897N200877B 4 30 Chapter 4 Preparation and Inspection ED580XT_E2 30_897N200877B indb 30 2018 09 20 15 50 23...
Страница 114: ...897N200877B 6 12 Chapter 6 Troubleshooting ED580XT_E2 30_897N200877B indb 12 2018 09 20 15 50 25...
Страница 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Страница 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...