897N200877B
6-2
Chapter 6 Troubleshooting
6.1 Troubleshooting
6.1.1 Problem with Displayed Images
Problem
Cause
Remedy
No images
The cart, monitor, processor or light
source is unplugged from the outlet.
Plug the cart, monitor, processor or light
source into the main outlet.
The cart, monitor, processor or light
source is OFF.
Power ON the cart, monitor, processor or
light source.
The endoscope is not connected
correctly to the light source.
Connect the endoscope properly to the light
source.
“4.6 Connecting Endoscope to Light
S o u r c e , P r o c e s s o r a n d R e l a t e d
Equipment”
Dark image *
1
The endoscope is not connected to
the light source correctly.
Connect the endoscope to the light source
correctly.
“4.6 Connecting Endoscope to Light
S o u r c e , P r o c e s s o r a n d R e l a t e d
Equipment”
The brightness level is set around
“MIN.”
Set the brightness level around 0.
Operation manual of the light source
The iris mode is set to “PEAK.”
Set iris mode to “AVE.”
Operation manual of the processor
Moisture or foreign matter (such
as dust, gauze fibers, metallic
fragments) adheres to the light
g u i d e c o v e r g l a s s o f t h e L G
connector.
Wipe off foreign matter on the light guide
cover glass of the LG connector with soft,
sterile gauze moistened with alcohol.
Moisture or foreign matter (such
as dust, gauze fibers, metallic
fragments) adheres to the objective
lens or light guide.
Wipe off foreign matter on the objective
lens or light guides with soft, sterile gauze
moistened with alcohol.
Halation
The brightness level is set around
“MAX.”
Set the brightness level around 0.
Operation manual of the light source
The iris mode is set to “AVE.”
Set iris mode to “PEAK.”
Operation manual of the processor
*1 When argon plasma coagulation (APC) is performed, some areas in the live image may become dark
due to luminous beam of argon plasma.
ED580XT_E2-30_897N200877B.indb 2
2018/09/20 15:50:25
Содержание ED-580XT
Страница 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Страница 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Страница 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Страница 35: ...897N200877B 2 3 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 3 2018 09 20 15 50 17...
Страница 48: ...897N200877B 2 16 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 16 2018 09 20 15 50 19...
Страница 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Страница 82: ...897N200877B 4 30 Chapter 4 Preparation and Inspection ED580XT_E2 30_897N200877B indb 30 2018 09 20 15 50 23...
Страница 114: ...897N200877B 6 12 Chapter 6 Troubleshooting ED580XT_E2 30_897N200877B indb 12 2018 09 20 15 50 25...
Страница 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Страница 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...