897N200877B
6-5
Chapter 6
Trouble
shooting
Problem
Cause
Remedy
Low suction
volume
T h e s u c t i o n v a l v e h a s b e e n
damaged.
Replace with a new reprocessed suction
valve.
T h e f o r c e p s v a l v e h a s b e e n
damaged.
Replace with a new forceps valve.
The suction tube is not attached
properly.
Reattach the suction tube.
The forceps valve is not attached
properly.
Attach the forceps valve properly and close
the lid.
Suction valve
does not return
to the original
position.
Solid materials or thick fluids have
adhered to the suction valve.
Detach the suction tube from the suction
unit. Detach the suction valve and replace
with a new reprocessed suction valve.
Suction valve is damaged.
Replace with a new suction valve.
Suction valve
cannot be
detached.
The suction valve or the control
portion of the endoscope has been
damaged.
Contact your local FUJIFILM dealer.
Fluid leaks from
forceps valve
during suction.
The forceps valve is not attached
correctly.
Attach the forceps valve properly and close
the lid.
The forceps valve is damaged.
Replace with a new forceps valve.
6.1.6 Problem with Related Equipment
♦
Problem with Endotherapy Devices
Problem
Cause
Remedy
Endotherapy
device cannot
be inserted.
The endotherapy device (such as
biopsy forceps) is left open.
Close the endotherapy device for insertion.
The handle of endotherapy device
(such as biopsy forceps) is held
firmly.
Loosen the grip to insert the endotherapy
device.
The endotherapy device has
difficulty being inserted due to
bending.
Reduce the angle of the bending section
slightly and then insert it.
The endotherapy device has an
abnormality.
Withdraw the endotherapy device and
replace it with a new one.
An endotherapy device which is
not compatible with this product is
used.
Use an endotherapy device compatible with
this product.
ED580XT_E2-30_897N200877B.indb 5
2018/09/20 15:50:25
Содержание ED-580XT
Страница 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Страница 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Страница 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Страница 35: ...897N200877B 2 3 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 3 2018 09 20 15 50 17...
Страница 48: ...897N200877B 2 16 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 16 2018 09 20 15 50 19...
Страница 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Страница 82: ...897N200877B 4 30 Chapter 4 Preparation and Inspection ED580XT_E2 30_897N200877B indb 30 2018 09 20 15 50 23...
Страница 114: ...897N200877B 6 12 Chapter 6 Troubleshooting ED580XT_E2 30_897N200877B indb 12 2018 09 20 15 50 25...
Страница 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Страница 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...