897N200877B
4-1
Chapter 4
Prepara
tion and Inspectio
n
Chapter
4
Preparation and
Inspection
This chapter describes the inspection and preparation methods to be performed before using
the endoscope, its accessories and related equipment.
Before using this product, perform preparation and inspection as per instructions provided
in this chapter. In addition, inspect related products used in combination with this product as
per instructions provided in respective operation manuals. If the inspection result shows any
abnormality, refer to “Chapter 6 Troubleshooting.” If the problem persists, or if any failure is
found, stop using the equipment and return it for repair according to “6.4 Returning Endoscope
for Repair.”
WARNING
• The entire surface and each channel of the endoscope and the accessories must be
reprocessed prior to first use, after any servicing, after any subsequent use, and after
storage as per instructions provided in the Reprocessing Manual, even if the accessories
were not used during a procedure. In addition, store this product as per instructions
provided in the Reprocessing Manual. Inadequate reprocessing or storage may cause
infection.
• Make sure to inspect the endoscope and accessories before use according to the
procedures provided in this manual. Do not use the equipment that shows any signs
of abnormality or irregularity. Use of abnormal equipment may lead to misdiagnosis or
increase risks to patient safety.
• During an inspection or procedure, use sterile water. If sterile water is not used, it can
create a risk of infection.
CAUTION
• Make sure to inspect the endoscope and accessories before use according to the
procedures provided in this manual. Do not use the equipment that shows any signs of
abnormality or irregularity. Use of abnormal equipment may cause equipment malfunction.
• If abnormalities and/or material changes including but not limited to cracking, flaking,
pitting, corrosion, etc. which can create rough surfaces or sharp edges, holes/depressions,
compromise sealed surfaces and/or negatively affect device functionality are found,
contact your local FUJIFILM dealer.
Note
The endoscope and accessories are not reprocessed before shipping from FUJIFILM.
Reprocess them according to the instructions given in the Reprocessing Manual
before using them in a procedure.
ED580XT_E2-30_897N200877B.indb 1
2018/09/20 15:50:19
Содержание ED-580XT
Страница 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Страница 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Страница 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Страница 35: ...897N200877B 2 3 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 3 2018 09 20 15 50 17...
Страница 48: ...897N200877B 2 16 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 16 2018 09 20 15 50 19...
Страница 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Страница 82: ...897N200877B 4 30 Chapter 4 Preparation and Inspection ED580XT_E2 30_897N200877B indb 30 2018 09 20 15 50 23...
Страница 114: ...897N200877B 6 12 Chapter 6 Troubleshooting ED580XT_E2 30_897N200877B indb 12 2018 09 20 15 50 25...
Страница 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Страница 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...