897N200877B
2-7
Chapter 2
Product Overview
♦
Control Portion
The control portion is used for operating each function of the endoscope. Hold this portion
during a procedure.
No.
Name
Function
10 Up/Down Angulation
Lock
Maintains the up/down angle of the bending section.
Moving this lock in the direction of F (Free) unlocks the up/
down movement of the bending section, allowing external
force to angulate the bending section freely.
Moving this lock in the direction opposite to F locks the up/
down movement and maintains the up/down angle of the
bending section.
11 Up/Down Angulation
Knob
Angulates the bending section upward or downward.
Turning this knob in the direction of U angulates the bending
section upward.
Turning this knob in the direction of D angulates the bending
section downward.
12 Left/Right Angulation
Lock
Maintains the right/left angle of the bending section.
Moving this lock in the direction of F (Free) unlocks the left/
right movement of the bending section, allowing external
force to angulate the bending section freely.
Moving this lock in the direction opposite to F locks the
right/left movement and maintains the right/left angle of the
bending section.
13 Left/Right Angulation
Knob
Angulates the bending section to the right or left.
Turning this knob in the direction of L angulates the bending
section to the left.
Turning this knob in the direction of R angulates the bending
section to the right.
14 Forceps Elevator Lever
Raises or lowers the forceps elevator.
Turning this lever in the direction of U raises the forceps
elevator.
Turning this lever in the direction opposite to U lowers the
forceps elevator.
15 Forceps Valve
One of the accessories for the endoscope. This valve is
attached to the instrument channel inlet to prevent leak or
backflow of air and/or fluids. In addition, an endotherapy
device is inserted into or a syringe is attached to this valve.
16 Instrument Channel
Inlet
Each endotherapy device is inserted from this inlet. An
endotherapy device or fluid injected with a syringe passes
through the instrument channel and comes out of the
instrument channel outlet in the distal end of endoscope.
17 Air/Water Valve Cylinder The air/water valve is attached to this cylinder.
ED580XT_E2-30_897N200877B.indb 7
2018/09/20 15:50:18
Содержание ED-580XT
Страница 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Страница 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Страница 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Страница 35: ...897N200877B 2 3 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 3 2018 09 20 15 50 17...
Страница 48: ...897N200877B 2 16 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 16 2018 09 20 15 50 19...
Страница 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Страница 82: ...897N200877B 4 30 Chapter 4 Preparation and Inspection ED580XT_E2 30_897N200877B indb 30 2018 09 20 15 50 23...
Страница 114: ...897N200877B 6 12 Chapter 6 Troubleshooting ED580XT_E2 30_897N200877B indb 12 2018 09 20 15 50 25...
Страница 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Страница 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...