897N200877B
1-13
Chapter 1
Precautions
CAUTION
• Do not use this product outside the operating environment specified in “Main
Specification.” Otherwise, it can cause product malfunction or failure.
• Do not apply unnatural force or impact on the insertion portion, bending section, control
portion, LG flexible portion, LG connector or video connector. Doing so may cause
malfunction of the endoscope.
• When holding the endoscope, hold it by the control portion. Handling it up by the insertion
portion or LG flexible portion is difficult to hold and may exert an unnatural force, resulting
in instrument failure.
• During an observation, do not perform close observation for an extended period of time.
Use the endoscope with a minimum necessary amount of brightness and time while
maintaining an appropriate distance. Thermal energy created by illumination may cause
burn injury.
• Immediately after detaching the LG connector from the light source, do not touch the light
guide prong with hands since it may be extremely hot. There is a risk of burn injury.
• If the brightness level of the light source or processor is high, the surface temperature at
and around the distal end of the endoscope may exceed 41°C. Do not allow the distal end
to remain in contact with the same site for an extended period of time. It may cause burn
injury.
• Do not use a mouthpiece that is damaged, deformed, or reveals other irregularities. Doing
so may cause injury in the oral cavity and/or equipment failure.
• Do not directly apply Xylocaine spray to the insertion portion. Do not use olive oil as a
lubricant for insertion. It may cause deterioration of the outer surface.
• Avoid forcible or excessive angulation as this may impose a severely heavy load on the
wire controlling the bending section. This may cause stretching or tearing of the wire.
• The lid of the forceps valve must be closed when using the endoscope. Not doing so can
reduce the efficacy of the endoscope’s suction system, making it impossible to perform
aspiration.
• Do not hold the bending portion tightly when attaching or detaching the distal end cap.
There is a risk of damaging the endoscope.
• Securely attach the distal end cap to the distal end of the endoscope before use. If the
distal end cap is not attached securely, the distal end cap may fall off during use.
• Use this product in combination with the distal end cap DC-07D. If another type of distal
end cap is used in combination with this product, the distal end cap may fall off during use.
ED580XT_E2-30_897N200877B.indb 13
2018/09/20 15:50:17
Содержание ED-580XT
Страница 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Страница 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Страница 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Страница 35: ...897N200877B 2 3 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 3 2018 09 20 15 50 17...
Страница 48: ...897N200877B 2 16 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 16 2018 09 20 15 50 19...
Страница 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Страница 82: ...897N200877B 4 30 Chapter 4 Preparation and Inspection ED580XT_E2 30_897N200877B indb 30 2018 09 20 15 50 23...
Страница 114: ...897N200877B 6 12 Chapter 6 Troubleshooting ED580XT_E2 30_897N200877B indb 12 2018 09 20 15 50 25...
Страница 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Страница 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...