897N200877B
Appendix-13
Appendix
<Electromagnetic Immunity Compliance Information and Guidance>
Table 6
Guidance and Manufacturer Declaration - Electromagnetic Immunity -
This device is intended for use in the following prescribed electromagnetic environments. Customers and users of
this product are advised to check that it is being used in such environments.
Immunity test
IEC 60601
Test level
Compliance level
Guidance
Electrostatic discharge
(ESD)
IEC/EN 61000-4-2
±2 kV contact
±4 kV contact
±6 kV contact
±2 kV air
±4 kV air
±8 kV air
±2 kV contact
±4 kV contact
±6 kV contact
±2 kV air
±4 kV air
±8 kV air
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast transient/
burst
IEC/EN 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
±2 kV for power supply
lines
±1 kV for input/output
lines
Main power quality should be
that of a typical commercial
or hospital environment.
Surge
IEC/EN 61000-4-5
±0.5 kV, ±1.0 kV
line to line
±0.5 kV, ±1.0 kV, ±2.0 kV
line to earth
±0.5 kV, ±1.0 kV
line to line
±0.5 kV, ±1.0 kV, ±2.0 kV
line to earth
Main power quality should be
that of a typical commercial
or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC/EN 61000-4-11
<5%
U
T
(>95% dip in
U
T
)
for 0.5 cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25 cycles
<5%
U
T
(>95% dip in
U
T
)
for 5 s
<5%
U
T
(>95% dip in
U
T
)
for 0.5 cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25 cycles
<5%
U
T
(>95% dip in
U
T
)
for 5 s
Main power quality
should be that of a typical
commercial or hospital
environment. If the user
of this product requires
continued operation during
power mains interruptions,
it is recommended that this
product is powered from an
uninterruptible power supply
or battery.
Power frequency
(50/60 Hz) magnetic field
IEC/EN 61000-4-8
3 A/m
3 A/m
The power frequency
magnetic field should
have the same level of
characteristics as a common
location in standard
business and hospital
environments.
Note
U
T
is the a.c. mains voltage prior to application of the test level.
ED580XT_E2-30_897N200877B.indb 13
2018/09/20 15:50:26
Содержание ED-580XT
Страница 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Страница 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Страница 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Страница 35: ...897N200877B 2 3 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 3 2018 09 20 15 50 17...
Страница 48: ...897N200877B 2 16 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 16 2018 09 20 15 50 19...
Страница 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Страница 82: ...897N200877B 4 30 Chapter 4 Preparation and Inspection ED580XT_E2 30_897N200877B indb 30 2018 09 20 15 50 23...
Страница 114: ...897N200877B 6 12 Chapter 6 Troubleshooting ED580XT_E2 30_897N200877B indb 12 2018 09 20 15 50 25...
Страница 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Страница 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...