CAUTION
• Use this product only in combination with related equipment described in this manual.
Otherwise, it creates a risk of equipment malfunction.
• Use the endoscopic CO
2
regulator described in this manual. If another insufflation device
is connected, the air/water supply function lessens and may result in improper cleaning of
the lens.
• Set the suction pressure between 40 and 53 kPa. If the suction pressure is too high,
the endoscope may adhere to mucous membrane, resulting in damage to the mucous
membrane.
• Do not grasp the bending section forcefully when attaching or detaching the distal end
cap. It may cause malfunction of the endoscope.
• Do not subject the video connector or the LG connector to strong impact. Install the
processor and the light source away from obstacles to prevent the video connector
connected to the processor and the LG connector connected to the light source from
accidental impact damage. During the operation of an electric bed, etc., ensure that
the video connector connected to the processor or the LG connector connected to the
light source does not hit the bed. Otherwise, the endoscope's video connector and LG
connector, processor and light source may malfunction.
• With regard to the amount of sterile water in the water tank, follow the instructions
provided in the operation manual of the water tank. If the amount of sterile water in the
water tank exceeds the limit, the air/water supply function may be disabled or it may cause
equipment failure due to contact with leaked sterile water.
• Attach the water tank to the specified position of the cart or light source. Otherwise, fluid
may leak from the connector of the water tank and it may come into contact with related
equipment, causing equipment failure.
• Firmly connect the suction tube from the suction unit to the suction connector on the LG
connector. If the suction tube is not attached properly, body fluid may drip from the tube
and come into contact with related equipment, causing equipment failure.
Note
• For details on how to use related equipment, refer to the operation manual of related
equipment.
• Before using this product, thoroughly read the operation manual of related equipment
used in combination with this product.
ED580XT_E2-30_897N200877B.indb 17
2018/09/20 15:50:17
Содержание ED-580XT
Страница 2: ...ii 897N200877B ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 15...
Страница 12: ...897N200877B 4 Introduction ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 16...
Страница 32: ...897N200877B 1 20 Chapter 1 Precautions ED580XT_E2 30_897N200877B indb 20 2018 09 20 15 50 17...
Страница 35: ...897N200877B 2 3 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 3 2018 09 20 15 50 17...
Страница 48: ...897N200877B 2 16 Chapter 2 Product Overview ED580XT_E2 30_897N200877B indb 16 2018 09 20 15 50 19...
Страница 52: ...897N200877B 3 4 Chapter 3 Workflow ED580XT_E2 30_897N200877B indb 4 2018 09 20 15 50 19...
Страница 82: ...897N200877B 4 30 Chapter 4 Preparation and Inspection ED580XT_E2 30_897N200877B indb 30 2018 09 20 15 50 23...
Страница 114: ...897N200877B 6 12 Chapter 6 Troubleshooting ED580XT_E2 30_897N200877B indb 12 2018 09 20 15 50 25...
Страница 116: ...897N200877B 7 2 Chapter 7 Service ED580XT_E2 30_897N200877B indb 2 2018 09 20 15 50 25...
Страница 134: ...897N200877B Appendix 18 Appendix ED580XT_E2 30_897N200877B indb 18 2018 09 20 15 50 27...