8 Connection
8.1 Transducers
PIOX S72*
2022-05-15, UMPIOX_S72xV1-9EN
74
8.1.3.3
Transducer cable with stainless steel conduit and stripped ends
• Remove the blind plug for the connection of the transducer cable.
• Insert the transducer cable into the junction box.
• Fix the transducer cable by tightening the cable gland.
• Connect the transducer cable to the terminals of the junction box.
Fig. 8.7:
Connection of the transducer cable with plastic jacket and stripped ends
1 – cap nut
2 – compression part
3 – basic part
4 – external shield, brushed back
Tab. 8.7:
Terminal assignment
terminal
connection
V
transducer (brown cable, marked white)
VS
transducer (red cable)
RS
transducer (red cable)
R
transducer (brown cable)
100 mm
20 mm
10 mm
4
cable gland
1
2
3