7 Mounting
PIOX S72*
7.2 Transducers
55
UMPIOX_S72xV1-9EN, 2022-05-15
Step 5: Installation of the transducers in the mounting fixture Variofix C
• Put coupling foil (or apply some coupling compound for a short-term installation) on the contact surface of the
transducers. The coupling foil can be fixed to the contact surface with a small amount of coupling compound.
• Position the transducers on the rail in such a way that the engravings on them form an arrow. The transducer cables
show in opposite directions (see Fig. 7.54).
• Adjust the transducer distance displayed by the transmitter (see Fig. 7.54).
• Slide the spring clips on the transducers (see Fig. 7.55).
• Fix the transducers by slightly tightening the tensioning screws. The end of each tensioning screw has to be placed
above the hole in the respective transducer (see Fig. 7.54 and Fig. 7.55).
• Correct the transducer distance, if necessary.
• Tighten the tensioning screws.
• Fix the spacing elements on the rail to mark the transducer position (see Fig. 7.54).
• Use the cable fixture to fix the transducer cables in order to protect them from mechanical strain (see Fig. 7.55).
• Put the cover on the rail (see Fig. 7.56).
• Tighten the screws (2) on both sides of the cover (see Fig. 7.56).
Notice!
If the signal is not sufficient for the measurement, use coupling compound instead of coupling foil. For information
concerning the coupling foil or coupling compound, see the safety data sheet.
Fig. 7.54:
Transducers in the rail (spring clip not shown)
1 – spacing element
2 – hole
a – transducer distance
Fig. 7.55:
Transducers in the rail
1 – equipotential bonding terminal
2 – spring clip
3 – tensioning screw
4 – cable fixture
1
2
a
2
4
3