18 Data logger
18.1 Configuration of the data logger
PIOX S72*
2022-05-15, UMPIOX_S72xV1-9EN
166
• Select in the scroll list a storage rate for storing the measured values in case the event occurs.
• Press ENTER.
• Enter the time interval for the measured values to be stored before the event occurs.
• Press ENTER.
• Enter the time interval for the measured values to be stored if the event is no longer activated.
• Press ENTER.
Storage rate
The storage rate is the frequency to transmit or store measured values. If a start time for storing the measured values is
defined, a storage rate has to be entered.
• Select the menu item
Configuration
.
• Press ENTER until the menu item
Storage rate
is displayed.
• Select a storage rate from the scroll list.
• Press ENTER.
• If
Customized
is selected, enter the storage rate.
• Press ENTER.
Ringbuffer
The data logger can be configured as linear logger or ringbuffer. If the ringbuffer is deactivated and the data logger is full,
the storing of measured values will be terminated. It can be continued after clearing the data logger. If the ringbuffer is
activated and the data logger is full, the oldest measured values will be overwritten. In ringbuffer mode, the remaining
capacity of the data logger is displayed during the measurement, e.g.:
Log
←→
: 1d 6h 57m
is displayed, if no measured values have been overwritten.
Log|
←→
| : 1d 6h 57m
is displayed, if the old measured values have been overwritten.
• Select the menu item
Configuration
.
• Press ENTER until the menu item
Ringbuffer
is displayed.
• Select
On
to activate the ringbuffer.
• Press ENTER.
If the ringbuffer is deactivated and the data logger is full, the storing of measured values will be terminated.
• Select
Off
to deactivate the ringbuffer.
• Press ENTER.
Special functions\Data logger\Configuration\Start storing\Event-based\
Storage rate (trig.)
Special functions\Data logger\Configuration\Start storing\Event-based\
Buffer time ->
П
Special functions\Data logger\Configuration\Start storing\Event-based\
Buffer time
П
->
Special functions\Data logger\Configuration\Storage rate
Special functions\Data logger\Configuration\Ringbuffer