7 Mounting
PIOX S72*
7.2 Transducers
61
UMPIOX_S72xV1-9EN, 2022-05-15
7.2.2.7
Mounting with tension strap and quick release clasp
• Cut the tension strap to length (pipe circumf at least 120 mm).
• Make sure that part (2) of the clasp is on top of part (1), see Fig. 7.67 a. The hooks of part (2) have to be on the outer
side of the clasp.
• Pull approx. 20 mm of the tension strap through the slot of the clasp to fix the clasp to the tension strap, see Fig. 7.67 b.
• Bend the end of the tension strap.
• Position the transducer on the pipe, see Fig. 7.68.
• Place the tension strap around the transducer and the pipe.
• Push the tension strap through part (2) and (1) of the clasp.
• Tighten the tension strap and engage it in the inner hook of the clasp.
• Tighten the screw of the clasp.
Caution!
The edge of the tension strap is very sharp.
Risk of injury.
→
Debur sharp edges.
→
Wear the required personal protective equipment.
→
Observe the applicable rules.
Fig. 7.67:
Clasp
Fig. 7.68:
Transducer on the pipe