7 Mounting
PIOX S72*
7.2 Transducers
45
UMPIOX_S72xV1-9EN, 2022-05-15
• Repeat the steps to fix the second rail, see Fig. 7.30.
Step 5: Installation of the transducers in the mounting fixture Variofix L
• Press the transducers firmly into their clamping fixtures in the covers until they are tightly fixed. The transducer cables
show in opposite directions (see Fig. 7.31).
• Adjust the transducer distance displayed by the transmitter (see Fig. 7.32).
• Fix the transducer cables with the strain relief clamp to protect them from mechanical strain, see Fig. 7.32.
• Put coupling foil (or apply some coupling compound for a short-term installation) on the contact surface of the
transducers. The coupling foil can be fixed to the contact surface with some coupling compound.
• Put the covers with the transducers on the rails.
• Correct the transducer distance, if necessary.
Fig. 7.30:
Pipe with 2 rails
Notice!
The arrows on the transducers and the covers have to point in the same direction.
Fig. 7.31:
Installation of the transducers in the covers
1 – cover
2 – transducer clamping fixture
1
2
1
2