10 Measurement
PIOX S72*
10.3 Measurement settings
111
UMPIOX_S72xV1-9EN, 2022-05-15
Depending on the selected source item, it is possible to output measured values, status values or event values.
10.3.6.1
Output of a measured value
• Select the list item
Options\Outputs\...\Values
.
• Press ENTER.
Output range
• Select a list item.
– 4...20 mA
– Other range
• Press ENTER.
If
Other range
is selected, enter the values
Output
MIN
and
Output MAX
.
The output range has to be > 10 % of the max. output value (
Output MAX
). If the output range is smaller, an error
message will be displayed.
Miscellaneous
Custom. Input 1
measured values of input quantities (e.g., temperature, density) which are
not used for calculation
In the menu item
Options\Assign inputs
it is possible to assign
configured inputs to customized inputs.
Custom. Input 2
Custom. Input 3
Custom. Input 4
Sound speed
Sound speed
sound speed
Tab. 10.2: Output of measured values, status values or event values
source item
measured value
event value
value
status
physical quantities
Flow quantities
x
x
Totalizers
x
x
Fluid properties
x
x
Miscellaneous
(Custom. Input 1...4)
x
x
Sound speed
x
x
Pulse
x
x
Diagnostic values
(except
Pig detection
)
x
events
Diagnostic values
\
Pig detection
x
Event trigger
x
Tab. 10.1: Configuration of the outputs
source item
list item
output
Source item
Values
Output range
Error value
Measuring range
Output info
Test signal
Test mea. range
see annex A
Options\Channel A\Outputs\...\Output
range